background image

SonarMite v3.0 (c)2009 Lymtech LLC 

 

Page 7 

6/13/2009 

warm  fresh  water  and  cold  salt  water  is  about  +/-30m/Second  which  represents  a 
change of depth of +/-2%, this must be viewed in context with typical depths of the 
survey,  if  a  typical  depth  is  10m  then  the  error  could  be  about  20cm. 

The  Sound 

Velocity  setting  is  the  largest  source  of  measurement  error  when  care  is  not 
taken in setting it correctly. 

Internal Battery Charging 

 

The  SonarMite  is  fitted  with  an  internal  sealed  Nickel  Metal  Hydride  battery  which 
provides 1.5Ahr @ 10vDC and a mains powered battery charger capable of bulk and 
float charging to give the most efficient battery cycle. The user should avoid running 
the SonarMite until the batteries are completely flat as there will be a high risk of both 
data  loss  and  permanent  damage  to  the  batteries.  The  charger  will  illuminate 
CONSTANT  RED  when  bulk  charging  and  FLASHING  YELLOW  when  charged, 
the charger illustrated may change if a local mains variant is required.  

DISCHARGE

CHARGE(on)

POWER(on)

READY(flashing)

CHARGE SEQUENCE

1

MEASURE BATTERY

CHARGE

POWER

READY

DISCHARGE

CHARGE(on)

POWER(on)

2

CHARGE

CHARGE

POWER

READY

DISCHARGE

POWER(on)

READY(flashing)

3

FINISHED

CHARGE

POWER

READY

DISCHARGE

POWER(on)

READY(flashing)

2

FINISHED

CHARGE

POWER

READY

DISCHARGE

DISCHARGE(on)

POWER(on)

1

DISCHARGE BATTERY

CHARGE

POWER

READY

DISCHARGE SEQUENCE

WARNING - Only the correct cha rger should be used with the Sona rMite

 

 

 

Содержание SONARMITE v3.0

Страница 1: ...irst to produce products in the field of instrumentation and software used within the Earth Science sectors Achievements to date include DGM the first digital ground modelling software to run a standa...

Страница 2: ...therproofing 8 DEPTH TRANSDUCER 9 Transducer Mounting 9 Avoid Stressing Transducers 10 BLUETOOTH INTERFACE 11 BluWave ID Password code 11 Bluetooth Serial Port Profile 11 Features 11 Bluetooth actual...

Страница 3: ...Blue Box system where the user interface is provided by integrated software running on a portable computer connected via a Bluetooth link The use of wireless technology enables the instrument to be w...

Страница 4: ...oncern in a survey sounder is a narrow beam width to prevent averaging of the returned signal The SonarMite uses Active Transducers these are digital as opposed to analogue components and use micropro...

Страница 5: ...arger Smart depth transducer c w 5m cable and embedded processor Serial data lead IP68 Transit case with DuPont Velco internal case The Trimble GeoXT is shown for illustration only The following is a...

Страница 6: ...types to create full XYZ hydrographic data using the SonarVista software package Total Station XYZ data plus Time XYZ data from RTKGPS plus Time XY data from DGPS plus Tide plus Time Range and Bearin...

Страница 7: ...e most efficient battery cycle The user should avoid running the SonarMite until the batteries are completely flat as there will be a high risk of both data loss and permanent damage to the batteries...

Страница 8: ...rm to the IP67 waterproofing standard this benchmark describes equipment which can withstand a light spray of water from a hose for a short period of time it does not imply the device is suitable for...

Страница 9: ...rtical adjustment Chemical and impact resistant plastic housing Designed to meet CE requirements Shielded piezoceramic element for noise free echo sounder display Standard cable length 7 6 m 25 Transo...

Страница 10: ...t mounting Avoid Stressing Transducers The Active Transducers include piezoelectric elements which are embedded within a resin compound If the user has to fabricate a custom mounting for the transduce...

Страница 11: ...assword code In some instances the local Bluetooth interface may request a password ID to set up pairing with the device for version 3 devices the pin number is always set to 1234 numeric one two thre...

Страница 12: ...obstructive walls and the type of walls the signal will need to go through and on the antenna that is fitted Assuming it is connected to a class 1 device then the maximum range should be between 50 an...

Страница 13: ...onnection This 4 way female connector is used to connect to either a PC serial connection or for an incoming ASCII data stream such as GPS in a Polled connection Transducer Charger connection This 4 w...

Страница 14: ...of configuration data held within its semi permanent FLASH memory area On running the instrument this data is copied into the current working RAM area of the processor Pressing any of the configurati...

Страница 15: ...lable in the SYS format Memory Operations R eset Pressing R Contol R will reset the SonarMite to its default values these values are saved in Flash memory and set in current working RAM memory only av...

Страница 16: ...Output Output Parameters Output message is eight numeric ASCII parameters floating format space delimited as id depth roll pitch heave battery qa flags cr lf where id The id number of the instrument...

Страница 17: ...DBT 1 6 f 0 48 M 5C SMDBT 1 6 f 0 48 M 5C SMDBT 1 6 f 0 48 M 5C Example NMEA DBT Output Format 3 DPT NMEA mode e g SMDPT 1 81 0 0 66 SMDPT 0 48 0 0 62 SMDPT 0 48 0 0 62 SMDPT 0 48 0 0 62 SMDPT 0 48 0...

Страница 18: ...0 116 1500 8 9 0 SYS 14 0 48 109 109 0 116 1500 8 9 0 SYS 14 0 48 109 109 0 116 1500 8 9 0 Example System Output Format 8 New SonarMite format e g 1 1 88 12 7 128 20 1 0 48 8 9 115 0 1 0 48 8 9 115 0...

Страница 19: ...und velocity setting for lighter fresh warm water only available in the SYS output mode This is a dangerous function and will change calibration values if saved C lear Reset output to format 0 used by...

Страница 20: ...SonarMite v3 0 c 2009 Lymtech LLC Page 20 6 13 2009 Typical Windows HyperTerminal settings Hyperterminal ASCII settings Hyperterminal Connection settings...

Страница 21: ...s have been changed to new NiMh giving greater duration 12Hrs and faster charge time External power cannot be used with these batteries The HPR function is now discontinued as most precision positioni...

Отзывы: