EN 62
7000 SERIES INDICATORS
Tare
A semi-automatic tare can be taken before the weighing cycle starts. Simply press the
TARE
key.
A preset tare value can be entered using the numeric keypad. Use the numeric keys to enter the preset tare value.
The digits will appear in line 2 of the lower part of the display, below the “Data:” prompt. When tare entry is
complete, press
TARE
to accept the value.
Discrete I/O
The Animal/Dynamic weighing application supports one new discrete input and two new discrete outputs.
The new input is a command for Start that can be issued when the Press
ENTER
to Start message is displayed.
The two new outputs are Working and Cycle Complete. The Working output turns ON when the averaging cycle is
in process and turns OFF after the average weight has been determined and is displayed. At this point, the Cycle
Complete output turns ON. The Cycle Complete output turns OFF when the next averaging cycle is started.
4.2.3
Configuration
The
FUNCTION
key
must be programmed for Animal/Dynamic weighing in setup at Application > Function
Key > Assignment in order for this application to operate and to access the setup steps for the Animal/Dynamic
weighing application. When the
FUNCTION
key is properly programmed, the Animal/Dynamic weighing branch of
setup is displayed as shown in Figure 4-1.
Figure 4-1: Animal/Dynamic weighing Application Configuration Menu
Operation
Mode
The Mode assignment is selected from a drop-down list. Options are:
Selection
Function
1
Application determines the average total weight on the
scale
2
Application determines the average total weight on the
scale and calculates the average weight per animal
The default value is 1.
Sampling Time
The Sampling Time setting determines the amount of time over which the scale will sample the weight on the scale
in order to determine an average total weight. The value is entered using the alphanumeric entry keys. Valid
settings are from 1 to 9.9 seconds, in 0.1 second increments. The default value is 5.0.
Auto Start
When enabled, Auto Start begins the weighing process when the weight on the scale exceeds a minimum weight
defined by the Start Threshold, and has reached stability. When the process is complete, the weight must fall below
20d (display divisions) and then exceed the threshold again to start the next cycle. This feature can be Enabled or
Disabled. By default, it is Disabled.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание T72XW AM
Страница 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...