7000 SERIES INDICATORS
EN 107
APPENDIX C COMMUNICATIONS
C.1 Serial Interface Parameters
One standard serial port and one optional serial port are supported with the T72XW indicator. They are designated
COM1 (standard port on the Main PCB) and COM2 (optional).
COM1
provides an RS-232 interface only. The RS-232 interface is a three-wire (TDX, RXD, and GND) with
selectable XON-XOFF handshaking.
Optional
COM2
provides RS-232 and RS-485 interfaces. The RS-232 interface is a three-wire (TDX, RXD, and
GND) with selectable XON-XOFF handshaking. The RS-485 connection is a two-wire interface but does not
support multi-drop communication with addressing. The port must be selected for use as either RS-232 or RS-485
due to the different operating requirements of the interfaces.
The baud rate can be configured from 300 to 115.2K baud and a checksum character can also be configured when
using one of the continuous output strings.
The T72XW indicator supports the following functions on the serial ports:
SICS (level 0 and level 1)
Reports
Continuous
The COM1 port is also used when flashing new firmware to the indicator.
Demand Output Mode
The demand output mode transmits data only when the indicator receives a print request. Print requests are sent to
the T72XW indicator when:
The operator presses PRINT
A discrete input selected as print is triggered
An ASCII “P” is sent through a demand or continuous port
Auto print is enabled and all conditions for auto print are met
When triggered, data is transmitted in a string programmed in the output template portion of setup. Demand mode
is used typically when sending data to a printer or PC on a transactional basis.
NOTE: When the demand output mode is assigned, the port is automatically set up to also receive certain ASCII
characters to duplicate keypad functions.
Output Templates
The T72XW indicator provides three programmable output templates to define the string of data to be transmitted.
Each of the templates is configured in setup and can contain up to 25 fields of data. The same template can be
sent to both COM1 and COM2 if both ports are programmed for demand output. It is also possible to send a
different data string from the two ports by assigning a different template to each one.
The default values for each of the templates follow:
Template 1
29.94 kg
10.32 kg T
19.62 kg N
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание T72XW AM
Страница 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...