![OHAUS T51XW Скачать руководство пользователя страница 1](http://html1.mh-extra.com/html/ohaus/t51xw/t51xw_quick-start-manual_3711909001.webp)
Quick Start Guide
T51 Indicators
Ohaus Corporation www.ohaus.com
i
T51 Indicators
Quick Start Guide
ASSEMBLY
Both the T51P and T51XW connect to a base scale through a Load Cell Cable that connects to
the Indicator’s Main Printed Circuit Board (PCB). Each model has a mounting bracket that can
be attached to a column mount bracket or a wall mount. To remove the mounting bracket,
remove the threaded knobs, holding the Indicator by hand.
CAUTION
:
ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE
INDICATOR BEFORE MAKING INTERNAL CONNECTIONS.
The process of opening the Housing varies slightly for each model:
T51P:
1. Remove the four Phillips head screws
from the rear housing.
2. Open the housing; disconnect the
ribbon cable from the top of the PCB.
T51XW:
1. Remove the four hex head screws from
the rear housing.
2. Open the housing by carefully pulling
the top of the front housing forward.
The T51’s Main PCB, with connectors, jumpers, terminals and switches is shown below.
Printed Circuit Board (PCB), T51 Indicator.
Jumper Connections:
For a 4-wire load cell with no sense wires, short Jumpers W1 and W2.
For a 6-wire load cell that includes sense wires, open Jumpers W1 and W2.
For load cells with an extra ground shield wire, connect the shield to the center position (GND).
+EXC +SIG -SIG -EXC
+SENS GND -SENS
Load Cell wiring.
Pin Connection
J4-1 +Excitation
J4-2 +Sense
J4-3 +Signal
J4-4 GND
J4-5 –Signal
J4-6 –Sense
J4-7 –Excitation
Open/shorted jumpers.
Power Supply
connector
Battery
connector J2
(T51P only)
Alternate
Load Cell
Terminal J4
Sense
Jumper W1
Sense
Jumper W2
Security Switch
SW2
RS232
Terminal J7
(T51XW only)
External Input
Terminal Block
COM2
Connector J8
(Optional:
RS232 or
RS422/485)
RS232 Pin Connector
(T51P only)
Recharge
option-T51P
(opposite side)
Load Cell
Connector Plug