Ohaus Corporation
Model:..............................................................
ALL
Service Training
Description:...............
Straingage OCL Adjustment
Sg_ocl
6.14.00
Page 1 of 1
Straingage Transducer OCL Adjustment
1.
Place the test weight in the center of the platform.
2.
Tare the balance.
3.
Move the weight to point A and record the reading.
4.
Move the weight to point B and record the reading.
5.
Move the weight to point C and record the reading.
6.
Move the weight to point D and record the reading.
•
If the reading at point A is negative, file at points 1 and 4 AT AN ANGLE.
•
If the reading at point B is negative, file at points 1 and 2 STRAIGHT ACROSS.
•
If the reading at point C is negative, file at points 2 and 3 AT AN ANGLE.
•
If the reading at point D is negative, file at points 3 and 4 STRAIGHT ACROSS.
♦
It is not recommended that you try to adjust more than –5 counts if the beam has been filed
already.
♦
If the beam has not been filed previously, you can adjust –10 counts.
♦
Remember, when filing you are weakening the beam. File a little at a time.