OHAUS Ranger 3000 Series Скачать руководство пользователя страница 1

 

SERVICE MANUAL 

RANGER

 

 

3000

 & 

RANGER

 

 

Count 

3000

 Scales 

 

 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 

Ohaus Corporation, 7 Campus Drive, Suite 310, Parsippany, NJ 07054 (973) 377-9000 

Содержание Ranger 3000 Series

Страница 1: ...SERVICE MANUAL RANGER 3000 RANGER Count 3000 Scales Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 973 377 9000...

Страница 2: ...t the time of publication but Ohaus Corporation assumes no liability arising from the use or misuse of this material Reproduction of this material is strictly prohibited Material in this manual is sub...

Страница 3: ...egal For Trade 1 9 1 7 1 Settings 1 3 1 7 2 Verification 1 4 1 7 3 Sealing 1 5 CHAPTER 2 TROUBLESHOOTING 2 1 Troubleshooting 2 1 2 2 Diagnostic Guide 2 1 2 2 1 Diagnosis 2 1 2 2 2 Checking Load Cells...

Страница 4: ...he Service Menu B 1 B 3 Capacity Menu B 2 B 4 Graduation Menu B 2 B 5 Service Span Calibration B 2 B 6 Service Linearity Calibration B 3 B 7 Ramp Menu B 4 B 8 Expand Menu B 4 B 9 End Menu B 4 LIST OF...

Страница 5: ...r 3000 Front Display with controls 1 6 1 3 LFT Lock Switch and Seal 1 9 2 1 Main Printed Curcuit Board 2 3 3 1 Screws securing Top Housing 3 2 3 2 Top and Bottom Bottom 3 2 3 3 Main Printed Circuit Bo...

Страница 6: ...serviceable components Appendix A Standard Calibration Explains procedures for Standard Calibration performed prior to using a scale and after service Appendix B Service Calibration Describes the Serv...

Страница 7: ...esolution 1 30000 1 30000 1 30000 1 30000 1 30000 LFT ON Resolution nMax 1 3000 1 3000 1 3000 1 3000 1 3000 Linearity 2d r0 0001 kg r0 0002 kg r0 0005 kg r0 001 kg r0 002 kg Repeatability 2d r0 0001 k...

Страница 8: ...ieces ID ID number input for library record edit recall APW Display Store an APW Secondary Function Long Press 2 Off Turn scale off Units Change the weighing unit Switch between Check Weigh Check Coun...

Страница 9: ...to battery operation if there is a power failure or the power cord is removed With AC power the scale is constantly charging so the battery charge indicator will remain lit until fully charged The sc...

Страница 10: ...oz lb oz End P r i n t C O M L O C k E n d Menu Items Reset Stable A Print Contnt Layout Data Tr End Reset Baud Parity Stop Handsh Alt Cm End L Cal L Setup L Read L Unit L Print L COM End 1 5 4 Menu...

Страница 11: ...gles through active weighing units Short Press in Menu Toggles through available settings Function Mode Back Short Press Selects function setting Long Press Selects active Mode Short Press in Menu ret...

Страница 12: ...ing The indicator will automatically turn off when the batteries are fully charged TABLE 1 6 Symbol Charge level 0 to 10 Remaining 11 to 40 Remaining 41 to 70 Remaining 71 to 100 Remaining LED models...

Страница 13: ...E n d Menu Items Reset kg g lb oz lb oz End Reset Stable A Print Contnt Layout Data Tr End Reset Baud Parity Stop Handsh Alt Cm End L Cal L Setup L Read L Mode L Unit L Print L COM End Note 1 only for...

Страница 14: ...not accessible Unit menu Auto Tare are locked at their current setting s Filter is locked at current setting Stable Range setting is locked at 1d Zero Range setting is locked at 2 Auto Zero Tracking...

Страница 15: ...ed with the most common cause first If the first remedy does not fix the problem proceed to the next remedy Before attempting to repair the balance read all chapters of this manual to be familiar with...

Страница 16: ...oad output The reading should meet the Load Cell specifications The Load Cell output should be very close to linear over its capacity range NOTE The following steps involve power applied to the scale...

Страница 17: ...capacities See Table 2 3 for typical readings Figure 2 1 Top view of main PCB NOTE Table 2 3 indicates typical readings Actual values can vary but should remain linear throughout the range If readings...

Страница 18: ...le on 3 Using a DVM measure the excitation voltage EXE and EXE in Figure 2 1 The reading should be 3 1 volts dc This is the excitation voltage for the Load Cell and is regulated If the voltage is lowe...

Страница 19: ...value 6 Adjust the Simulator to simulate 0 load 50 load and 100 load for the capacity that the scale is rated for See Table 2 3 for values to use If the resulting readings are unstable the Main PC Boa...

Страница 20: ...elow Power On Zero limit Err 8 3 Over Range Error Weight reading exceeds Overload limit Err 8 4 Under Range Error Weight reading below Underload limit Err 8 5 Tare out of range Adjust tare value to be...

Страница 21: ...to stabilize 3 1 1 Preventive Maintenance Checklist The scale should be inspected and checked regularly as follows 1 Remove the Pan and Sub Pan to inspect and clean the area beneath the Pan 2 Clean th...

Страница 22: ...le over Remove the eight screws holding the Housing in place See Figure 3 1 3 Move the top housing a little bit backward and then raise and remove the top housing 4 Separate Top Housing from Bottom Ho...

Страница 23: ...he PCB to the bottom housing 3 If either the Main PCB or the Load Cell is to be replaced note the order of the wire colors and then disconnect the Load Cell Cable from the Main PCB See Figure 3 2 4 Po...

Страница 24: ...nd washers that hold the Load Cell assembly frame and load cell to the Bottom Housing Remove the bolts and their washer pairs that hold the Load Cell Frame to the Load Cell Use a high leverage Allen w...

Страница 25: ...edure requires test masses equal to the scale s capacity See Table 3 2 There are four Overload Stop Bolts See Figure 3 5 Adjust the Overload Stops per Table 3 2 1 Adjust the Overload Stop Bolt so that...

Страница 26: ...SETTINGS Model Max Capacity Overload Stop Gap R3000 RC 3000 Scale Load Cell A B mm C D mm R31P E 1502 RC31P1502 1 5kg 3kg 0 6 0 5 R31P E 3 RC31P3 3kg 5kg 0 8 0 7 R31P E 6 RC31P6 6kg 10kg 1 5 1 1 R31P...

Страница 27: ...battery leads 2 Pull the quick connect tabs free of the battery posts and lift out the battery Figure 3 7 Rechargeable battery 3 Position the replacement battery in the compartment on an angle suffici...

Страница 28: ...required to test the Ohaus Ranger 3000 Ranger Count 3000 scales must meet the requirements of ASTM Class 4 or OIML F2 Tolerance The mass values are listed in Table 4 1 TABLE 4 1 TEST MASS VALUES Model...

Страница 29: ...ch the specification for each model listed in Table 1 1 1 Power on the balance The reading on the display should be 0g 2 Select a mass weighing near the maximum capacity of the balance and place it on...

Страница 30: ...lity test set up the instrument as follows Set Up Follow the steps in Appendix A Section A 2 Setup and Calibration Record Settings Zero Tracking Setting ____________ Displayed Units ____________ Mass...

Страница 31: ...verage 6 Mean Reading 1 Reading 2 Reading 3 Reading 4 Reading 5 7 Reading 6 Reading 7 Reading 8 Reading 9 Reading 10 10 Mean ________ 6 Calculate the Delta for each reading and record in the work shee...

Страница 32: ...and press ON Zero Off 3 Place the test mass on the Scale record the weight and remove 4 Place Load 2 on the Scale and press ON Zero Off 5 Place the test mass on the Scale record the weight and remove...

Страница 33: ...1 Place the test weight in the center of the Weighing Pan 2 Tare the balance 3 Move the weight to point A and record the reading 4 Move the weight to point B and record the reading 5 Move the weight...

Страница 34: ...le in the field NOTE In all cases where a part is replaced the scale must be thoroughly checked after the replacement is made The scale MUST meet the parameters of all applicable specifications in thi...

Страница 35: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 2 Ranger Count 3000 Series Service Manual 5 1 Ranger 3000 SCALES PARTS Figure 5 1 Ranger 3000 Scales Parts...

Страница 36: ...1PE15JP 30037405 LoadCell AMI C3 40kg R31P30 R31PE30JP 7 30037435 Spider Down R31 R21 RC31 RC21 V71 8 30037420 Hardware Kit R31 R21 RC31 RC21 V71 9 30037398 Housing Bottom R31 10 30037421 Feet R31 R21...

Страница 37: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 4 Ranger Count 3000 Series Service Manual 5 2 RANGER COUNT 3000 SCALES PARTS Figure 5 2 Ranger Count 3000 Scales Parts...

Страница 38: ...I C3 40kg RC31P30 7 30037435 Spider Down R31 R21 RC31 RC21 V71 8 30037420 Hardware Kit R31 R21 RC31 RC21 V71 9 30037398 Housing Bottom R31 10 30037421 Feet R31 R21 RC31 RC21 V71 11 30037422 Cover Opti...

Страница 39: ...alibration process 2 Allow the scale to warm up for approximately five minutes after stabilizing to room temperature 3 To abort calibration press Exit key or turn the scale off anytime during the cali...

Страница 40: ...accept SpaN will then be shown Press Yes to begin the span calibration 0 kg will be displayed SpaN 0 3 Press Yes to accept C will be displayed while zero reading is stored Next the display shows the...

Страница 41: ...uch the scale or the surface it rests on while calibration is in progress as it will cause the process to fail B 1 Service Menu Structure B 2 Entering the Service Menu Turn the scale off Ranger Count...

Страница 42: ...e Service Span Calibration menu Set the graduation of the scale 0 00005 0 02 Grad B 5 Service Span Calibration When SPAN appears press YES to accept the Service Span Calibration menu or No to advance...

Страница 43: ...to indicate the calibration unit The example shows the mid calibration point value for a Scale Capacity of 3 kg Place the specified calibration weight on the pan and press Yes C flashes while the mid...

Страница 44: ...o the Expand menu item Note Ramp is used to troubleshoot load cell problems 0 represents no signal from the A D 100 reflects the upper signal range of the A D ramMp 0 0 to 100 0 B 8 Expand Menu When e...

Отзывы: