 >0<  On/Off/Zero

Press this key to turn the scale on. This key is also used to rezero
the display. To turn the scale off, press and hold this key until the

display indicates OFF, then release.

Unit/Cal

Press this key to change the weighing unit. Press and hold to
begin calibration.

Weighing

Slide the pan/cover to Weighing until it locks into position. Place
the scale on a firm level surface. Turn the scale on. When the  *
symbol is displayed the scale is ready for use. Place sample on

the pan and read the displayed result. The * indicator appears
when the reading is stable. If using a container, press 

>0<

 to tare

the container before adding material.  If the capacity of the scale is
exceeded, an E will appear on the display and the load should be
removed immediately.

Calibration

Refer to Specifications for the correct calibration weights.

1. Press and hold 

Unit/Cal

 until “CAL” is displayed, then

release it.

2. Press

 >0<

. The required weight will appear in grams.

3. Place the appropriate weight on the pan, then press 

>0<

4. “PASS” will be displayed when calibration is complete.

Note

::::: If “CAL E” is displayed a calibration

error has occurred (for example, an in-
correct weight was used or the pan was
disturbed). Repeat the procedure.

Operation

This is a precision weighing in-

strument, handle with care.
Slide the pan/cover as shown to
expose the display and keys.
The pan/cover can slide to ei-
ther Weighing   or Storage 
positions.

NOTE

: For Weighing, slide pan/

cover all the way until a click is
heard (pan locks into weighing
position).  Do not apply exces-
sive force when sliding the pan,

and do not press on the pan or
try to remove it.

Once pan is locked into Weighing position, the scale is ready for
use.  Batteries are already installed.

To access the batteries, slide the pan/cover
to battery position   as shown and press
the release tab.  Install battery with (+)
facing up.

     

Specifications

 

Model

         PS121T

         PS251T

 Capacity (g)

120

250

 Readability (g)

0.1

0.1

 Calibration weight

100g

200g

 Weighing Units

   g, thk, tsg, trc

 Tare range

    to capacity by subtraction

 Stabilization time

     2 seconds

 Optimum operating temp.

65° - 77°F/18° - 25°C

 Power requirements

2-CR2032 Lithium batteries (included)

 Auto Shut-off

2 minutes no activity

  Overload  indication

          [E]

Register your scale under 

Warranty Registration at www.ohaus.com