Guardian
TM
Hotplate-Stirrers
EN-19
4.4. Using the Single Point Calibration (SPC) Feature
4.4.1. Start a Single Point Calibration (SPC)
1.
To enter the
Single Point Calibration
, the unit must first be in the top level of
the settings menu. Please refer to
Accessing / Exiting
Section for how to
enter the top level menu. Then wait for the screen to show
CAL
.
2.
Briefly press the left knob to enter the
Calibration
settings menu.
3.
Briefly press the left knob to enter the
SPC
settings menu.
4.
Briefly press the left knob again to change the calibration temperature.
The temperature will begin to blink to indicate that it can be modified.
5.
Rotate the left knob to scroll to the desired temperature.