SERVICE MENU 23
Page 23 of 43
SERVICE MANUAL FC5718/R
© OHAUS Corporation
Here you can read off the revisions number of the control board. From the revison number 2 and
the software version 1.33 on or higher, you have the possibility to check the function of the
imbalance sensor.
• Activate the operation menu as described under „Activation of the operation
menu (Part 1)”.
• Turn the potentiometer, until the special digit „d“ appears in the service
display.
• Example: Revision number 0
Revision number 1
without automatical
with automatical
check
check (in this case (off))
2.2 Activation of the Service menu (Part 2)
By pressing the key stop for longer than one second you will be asked for a code in
the display „rpm/rcf“. Adjust the code to “
9876
“ by the potentiometer and press once
again the key stop.
Now you can carry out the below shown adjustments
2.2.1 Submenu Adjustment of the centrifuge type
Here you can adjust the different centrifuge types that are stored on the
control board.
• Activate the operation menu as described under „Activation of the operation
menu (Part 1)”.
• Activate the service menu as described under „Activation of the service
menu (Part 2)”.
• Turn the potentiometer, until the special digit „t“ appears in the service
display.