Defender 6000 Indicators
EN-27
6. The display shows
--C--
, and then
-DONE-
when the calibration is finished.
7. Then the display shows
Lin
. Press the
Exit
button to exit.
Note:
If calibration is failed,
CAL E
is displayed for 3 seconds and the previous calibration data is restored.
The scale exits to the active weighing mode and displays the actual weight value in the current
weighing unit.
If after waiting for 40s the calibration is still not successful,
CAL E
is displayed for 3 seconds and the
previous calibration data is restored. The scale exits to the active weighing mode and displays the
actual weight value in the current weighing unit.
4.2.4 Linearity Calibration
[
LIN
]
Linearity calibration uses 3 calibration points. The full calibration point is established with a weight on the scale.
The mid calibration point is established with a weight equal to half of the full calibration weight on the scale. The
zero calibration point is established with no weight on the scale. The full calibration and mid calibration points
can be altered by the user during the calibration procedure.
Calibration procedures:
1. Long press the
Menu
button until you see
C.A.L
. Press the
Yes
button.
2. Short press the
No
button several times to navigate until you see
LIN
.
Press the
Yes
button.
3. The display flashes with
0
kg and the calibration unit. With no weight on the pan, press the
Yes
button to
establish the zero point.
4. The display shows
--C--
, and then moves to flash with the first calibration point and calibration unit based
on the capacity and unit you set in the capacity menu. (For example,
015.000
kg). If you do not need to
change the calibration point, skip to step 6.
5. To change the calibration point:
i-DT61PW: short press the
No
button several times until the desired digit appears. Short press the
Yes
button to accept the digit and move to the next one. Repeat the process until all the digits are correct.
Press the
Yes
button to accept the calibration point. The display flashes with the calibration point you
set.
i-DT61XWE: input the calibration point through the numeric keys. (Do not press the
Yes
button here
until you finish step 6).
6. Place a calibration mass of the specified weight on the pan and press the
Yes
button.
7. The display shows
--C--
, and then moves to flash with the second calibration point and calibration unit
based on the capacity and unit you set in the capacity menu. (For example,
030.000
kg).
Note:
If after waiting for 40s the calibration is still not successful,
CAL E
is displayed for 3 seconds and the
previous calibration data is restored. The scale exits to the active weighing mode and displays the actual
weight value in the currently selected weighing unit.
8. Repeat step 5 and 6.
9. The display shows
--C--
, and then
-DONE-
when the Linearity calibration is finished.
10. Then the display shows
GEO
. Press the
Exit
button to exit.
4.2.5 GEO Adjustment
[
GEO
]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current location. Settings
from 0 to 31 are available with 12 being the default.
Please refer to
the
Table of Geo Values
section in the
Technical Data
chapter to determine the GEO factor
that corresponds to your location.
To set the GEO factor:
1. Long press the
Menu
button until you see
C.A.L
. Press the
Yes
button.
2. Short press the
No
button several times to navigate until you see
GEO
.
Press the
Yes
button.
3. The display flashes with the Geo point (For example,
12
).
4. Short press the
No
button several times until the desired GEO number appears. Press the
Yes
button to
finish setting.
5. Then the display shows
C.test
. Press the
Exit
button to exit.
Содержание Defender 6000