EN-10
Defender 3000 Indicators
3. MENU SETTINGS
The User Menu allows the customizing of scale settings.
Note
: additional sub-menus (USB, Ethernet and Print2 menus in the table below) may be available if interface
options are installed. See interface user manual for the additional setting information.
3.1.
Menu Navigation
3.1.1
User Menu (in segments)
C.A.L S.E.t.U.P r.E.A.d mM.O.d.E U.n.I.t R.s.2.3.2 P.r.I.N.t U.S.B E.t.H.N.E.t P.r.I.N.t2 L.o.c.k E.n.d
ZErO
SpaN
LINE
GEO
C.test
End
reset
C.UNIt
rANGE
CAP1
Grad1
CAP2
Grad2
P.ZerO
P.UNIt
A.tare
End
reset
Stable
ZErO
FIltEr
AZt
B.LIGHt
b.COLOr
SCreeN
A.Off
P.SAuEr
CEC
End
reset
wWeIGH
COUNt
CHECK
tOtAL
tABLE
End
reset
kg
g
lb
oz
lb:oz
t
End
reset
BAUd
PArItY
StOP
H.SHAKE
ALt.P
ALt.t
ALt.Z
End
reset
ASSIGN
Stable
mMODE
tImME
lINkto
C.SUmM
tEmMp
End
reset
BAUd
PArItY
StOP
H.SHAKE
ALt.P
ALt.t
ALt.Z
ENd
reset
IP.Addr
u.E.r.
ENd
reset
ASSIGN
Stable
mMODE
tImME
lINkto
C.SUmM
tEmMp
ENd
reset
L.ALL
L.OFF
L.ZErO
L.PrINt
L.UNIt
L.mModE
L.mMeNU
L.tArE
End
Notes:
Some modes/units may not be available in all models.
When LEGAL FOR TRADE is turned
ON
(the security switch is in the locked position), the menu settings will be
affected as below:
Calibration function is disabled.
Zero Range setting is locked at 2%.
Stable Range setting is locked at 1d.
Auto-Zero Tracking setting is locked at 0.5d.
Filter and Units are locked at their current settings.
Stable Only is locked at On.
Lb
:
oz unit is locked at Off.
Power saving mode is disabled.
4. MAINTENANCE
CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.
4.1.
Cleaning
For i-DT33P, the housing may be cleaned with a cloth dampened with a mild detergent if necessary.
For i-DT33XW, use approved cleaning solutions for the stainless-steel Indicator housing and rinse with water.
Dry thoroughly.
WARNING
: Electric Shock Hazard. Disconnect the equipment from the power supply
before cleaning.
Make sure that no liquid enters the interior of the instrument.
Attention:
Do not use solvents, harsh chemicals, ammonia or abrasive cleaning agents.