Defender 3000 Indicators
EN-29
c)
Repeat the process until all the digits are correct.
d)
Press the
Yes
button to accept calibration point. It is flashing on the display.
6.
Place a calibration mass of the specified weight on the pan and press the
Yes
button.
7.
The display shows
--C--
, and then
-DONE-
when the calibration is finished.
8.
Then the display shows
Lin
. Press the
Exit
button to exit.
Note:
If calibration is failed,
CAL E
is displayed for 3 seconds and the previous calibration data is restored.
If after waiting for 40 seconds the calibration is still not successful,
CAL E
is displayed for 3 seconds
and the previous calibration data is restored.
4.2.4 Linearity Calibration
[
LIN
]
Linearity calibration uses 3 calibration points. The full calibration point is established with a weight on the pan.
The mid calibration point is established with a weight equal to half of the full calibration weight on the pan. The
zero calibration point is established with no weight on the pan. The full calibration and mid calibration points
can be altered by users during the calibration procedure.
Calibration procedures:
1.
Long press the
Menu
button until you see
mM.E.N.U
.
2.
Release the button and wait for the display to show
C.A.L
. Press the
Yes
button.
3.
Short press the
No
button several times to navigate until you see
LIN
.
Press the
Yes
button.
4.
0
kg and the calibration unit are flashing on the display. With no weight on the pan, press the
Yes
button to
establish the zero point. The display shows
--C--
.
5.
Then the first calibration point and the unit are flashing on the display based on the scale capacity and unit
you set in the setup menu. (For example,
015.000
kg). If you do not need to change the calibration point,
skip to step 7.
6.
To change the calibration point:
a)
Short press the
No
button several times until the desired number appears.
Note
: press the
Back
button can decrease the digit.
b)
Short press the
Yes
button to accept the number and move on to the next digit.
c)
Repeat the process until all the digits are correct.
d)
Press the
Yes
button to accept calibration point. It is flashing on the display.
7.
Place a calibration mass of the specified weight on the pan and press the
Yes
button. The display shows
--C--
.
8.
The second calibration point and the unit are flashing on the display based on the capacity and unit you
set in the setup menu. (For example,
030.000
kg)
Note:
If after waiting for 40s the calibration is still not successful,
CAL E
is displayed for 3 seconds and the
previous calibration data is restored.
9.
Repeat step 6 and 7.
10. The display shows
--C--
, and then
-DONE-
when the Linearity calibration is finished.
11. After that the display shows
GEO
. Press the
Exit
button to exit.
4.2.5 GEO Code Adjustment
[
GEO
]
Geographical Adjustment Factor (GEO) code is used to adjust the calibration based on the current location.
Settings from 0 to 31 are available with 12 being the default.
Please refer to
the
Table of Geo Code Values
section in the
Technical Data
chapter to determine the GEO
factor that corresponds to the indicator's location.
To set the GEO factor:
1.
Long press the
Menu
button until you see
mM.E.N.U
.
2.
Release the button and wait for the display to show
C.A.L
. Press the
Yes
button.
3.
Short press the
No
button several times to navigate until you see
GEO
.
Press the
Yes
button.
4.
The Geo point is flashing on the display (For example,
12
).