Defender 3000 Indicators
EN-21
3.4 Check
Use this application to compare the weight of items to a target weight range.
The display color will change according to the comparing result:
Red indicates over the target weight range
Green indicates within the target weight range.
Yellow indicates under the target weight range.
3.4.1 Set Check Limits
1.
Press and hold the
Mode
button until
CHECK
is displayed.
2.
When the
Mode
button is released, the display shows
CLr.CHK
.
If you want to use the stored under and over value of the last time, press the
No
button, and the scale
enters check mode directly.
If you want to set new over and under values, press the
Yes
button and continue to the next step.
3.
The display shows
UNDEr
. Press the
Yes
button to edit the under value.
4.
000000
is flashing on the display with the unit you set.
5.
To set a new under value:
a)
Short press the
No
button several times until the desired number appears.
Note
:
Press the
Back
button can decrease the digit.
If you need to set a negative value, press the
Back
button when the first digit is zero or press the
No
button when the first digit is nine.
b)
Short press the
Yes
button to accept the number and move to the next digit.
c)
Repeat the process until all the digits are correct.
d)
Press the
Yes
button to accept the value. Then the display will show
OuEr
.
Note
: Please refer to the
Button Navigation
section for details of setting digits.
6.
Repeat step 3 to 5 to set the over value.
7.
If the values you set are invalid, the display will show
--NO--
and go back to reset the Under and Over
value.
8.
If the values you set are valid, the scale will enter check mode.
Note
: Short press the
Function
button can display the over and under value.
3.4.2 Positive Check
Positive check is used to determine when the material added to the scale is within the target weight range. In
this case the under and over values must be positive values. (The over value must be greater than the under
value.)
To start, add material to the scale pan until the display turns green.
3.4.3 Negative Check
Negative check is used to determine when the material removed from the scale is within the target weight
range. In this case the under and over values are both negative values. The under value must be greater than
the over value. (For example: the under value is -10; the over value is -15).
To start, place a package or bin of material on the scale and press the
Tare
button.
Remove a portion of the package or material until the display turns green.
To continue weighing multiple portions off the scale's pan, tare the scale between each portion.
3.4.4 Zero Check
Zero check is used when comparing subsequent samples to an initial reference sample. In this case, the under
value must be zero or a negative value and the over value must be zero or a positive one.
Place the reference sample on the scale and press the
Tare
button.
Remove material from the scale pan until the display turns green.
3.4.5 Application Settings
The application can be customized for user preferences. Please refer to
Application Settings
section in
Weighing Mode for details about how to enter application settings.