Courier 7000
EN-55
11.3 OHAUS Commands
Commands listed in the following table will be acknowledged by the scale.
The scale will return “ES” for invalid commands.
Please add
\r\n
after each command to send
(
\r
refers to
\return
;
\n
refers to \newline)
OHAUS Commands
Command
Function
P
Print displayed weight (stable or unstable).
IP
Print displayed weight immediately (stable or unstable).
CP
Print weight continuously.
SP(0-60)
Print weight when stable.
Z
Same as pressing Zero Key.
T
Same as pressing Tare Key.
TA
Set/inquiry of tare weight value.
•
Set: TA
tare value unit
Example: TA 5 kg
•
Inquiry: TA
U
Set/inquiry current display unit.
•
Set: U
unit ID
Example: U 1
Please check the following table 11-5 for the ID of each unit.
•
Inquiry: U
M
Set/inquiry current application mode.
•
Set: M application
ID
Example: M 0
Please check the following table 11-6 for the ID of each application.
•
Inquiry: M
PSN
Print terminal's serial number.
PV
Print name, software revision and LFT ON (if LFT security switch is turned ON).
H x "text"
Enter print output string, x = string number (1-5), "text" = string text up to 40 alphanumeric
characters.
F x "text"
Enter Footer String content, x = string number (1-2), "text" = string text up to 40
alphanumeric characters.
NOTE
:
•
The italic content in each command needs to be filled with actual value.
•
Space in each command is needed. Please pay attention to it when input commands.
Unit ID
Unit Name
Abbreviation
0
Gram
g
1
Kilogram
kg
7
Pound
lb
8
Ounce
oz
20
Pound:Ounces
lb:oz
Application ID
Unit Name
0
Weighing
4
Dynamic
6
Totalization
18
Weight Alert
Table 11-5 Unit ID Table 11-6 Application ID