11
How can faulty calibration be resolved?
Faulty calibration can only be resolved by a precise re-calibration. Where the conditions for
precise calibration are lacking, this can be mitigated by a reset to factory settings. This function
gives the user the option to reset the sensors to their initial position, however, this will not
correspond to the current behaviour status of the sensors.
Open the
sub-menu “Calibration”
with
[MENU]
and press
[OK].
The relevant temperature value can be calibrated on the sensors for oxygen and pH. Please bear
in mind that the normal comparative value used must have a very high quality of <0.1 °C, and
the sensor must be exposed to a very constant temperature for 10 minutes. To avoid extreme
changes, the temperature range can only be changed by ± 1°C.
11.1.
Temperature pH / O2
The calibration can be aborted at any time using
the
[◄]
key.
1. Open the menu selection using
[MENU].
2. Use
[
▲
]
or
[
▼
]
for calibration and confirm with
[OK].
3. Use
[
▲
]
or
[
▼
]
to change to “temperature” of the
selected sensor and confirm with
[OK]
.
4. “actual temperature” and “target temperature” are
shown. Use
[
▲
]
or
[
▼
]
to select the desired tempera-
ture and confirm with
[OK].
[OK]
[OK]
Temperature calibration of the
pH sensor 1430817000
actual temperature
25.1 °C
target temperature
25.1 °C
modif. possible by ± 1 °C
raise the value with
▲
lower the value with
▼
save with button
[OK]
interrupt with button
[◄]
Main menu
calibration
pH redox sensor
pH redox value
temperature
oxygen sensor
oxygen value
temperature
Main menu
calibration
temperature
[
▲
]
or
[
▼
]
Содержание udox
Страница 1: ...dox Operating instructions Version 1 6 8...
Страница 2: ......