Aging Cells

#175-25 Stainless Steel Grade 303, 260 mL

#175-30 Stainless Steel Grade 303, 500 mL

#175-50 Stainless Steel Grade 316, 500 mL

Instruction Manual

Updated 6/14/2019

Ver. 4

OFI Testing Equipment, Inc.

11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.

Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com

©

Copyright OFITE 2014

Содержание 175-25

Страница 1: ...Stainless Steel Grade 303 500 mL 175 50 Stainless Steel Grade 316 500 mL Instruction Manual Updated 6 14 2019 Ver 4 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 3...

Страница 2: ...lecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Components 3 Parts and Accessories 4 Specifications 5 Operation 6 Maintenance 8 Warranty and Return Policy 9 T...

Страница 3: ...perature testing up to 500 F 260 C For prolonged exposure to elevated salinity at high temperatures such as 20 000 mg L chlorides at 350 F 176 6 C cells con structed from premium metals such as Hastel...

Страница 4: ...g Cell Body For Temperatures Below 400 F Qty 3 175 14 Set Screw Diameter Qty 3 175 15 Wrench for Set Screw 175 16 Valve Stem 175 47 O ring for Outside of Aging Cell Body Viton 90 Qty 2 175 30 SP Spare...

Страница 5: ...17 Viton O ring for Valve Stem 175 09 2 Viton O ring for Inside of Aging Cell Body 175 47 Viton O ring for Outside of Aging Cell 175 62 Viton O ring for Teflon Liner Plug 175 63 Viton O ring for Teflo...

Страница 6: ...e outer perimeter on the top and bottom of the cells Failure to do so can damage the rollers in the oven Teflon 175 46 Viton 175 47 and Buna N 175 54 o rings are available Cell Corrosion Test fluids u...

Страница 7: ...ill be tested refer to the chart on page 5 Place the correct volume of fluid into the cell Be careful not to get fluid in the o ring groove or on the o ring Do not overfill the cell 3 Clean any spille...

Страница 8: ...ell from the oven and allow it to air cool until the sample temperature is 300 F 149 C or less The cell may then be either air or water cooled The sample temperature must be equal to the ambient tempe...

Страница 9: ...e to the threads on the outer cap and cell body Every three years yearly for highly corrosive conditions 1 Sand blast the metal components cell body inner cap outer cap 2 Visually inspect the cell bod...

Страница 10: ...ny losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate hea...