A packing set is included with
TURBOVAC
models with a “C” in the type desi-
gnation. Use this packing set after detaching the pump from the system.
Faulty (leaky) packing of a
TURBOVAC
will nullify the guarantee.
Ordering data
Packing set for high-vacuum connection flange
Order No.
DN 100/160
200 91 240
DN 200
200 91 295
DN 250, 6” ANSI
200 91 262
Pack the pump so that it cannot be damaged during shipping and so that no
contaminants can escape from the packaging. Protect in particular the flan-
ges, the coolant connection nipples and the cable grommets.
If you return a pump to Oerlikon Leybold Vacuum, be absolutely sure to
observe the instructions given in Section 5.2.
Maintenance / Troubleshooting
31
GA05118_1502 - 12/2006 - © Oerlikon Leybold Vacuum
6
Troubleshooting
When the connector cable to the
TURBOVAC
is attached, the outputs of
the
TURBOTRONIK
frequency converter are not free of voltage.
Before commencing troubleshooting procedures, make the following simple
checks:
■
Is the TURBOVAC being supplied with electrical energy?
■
Are the connections . . .
- from the mains power cord to the frequency converter
- at the connector cable from the frequency converter to the mains net-
work in good working order?
■
If a water flow monitoring device is connected, is it functioning properly?
Check the water flow monitoring device by jumping its terminals and star-
ting the
TURBOVAC
.
■
Is the forevacuum pressure sufficient?
■
Is the vacuum chamber free of leaks?
Observe also the troubleshooting instructions for the
TURBOTRONIK
.
Warning