Computer interface
GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
83
6.3.36
TLC
Torr lock. See Chapter 5.4.4 Torr lock (tr-L),
S: TLC[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>
6.3.37
TRA
Test RAM. Test the RAM. See Chapter 5.4.6 RAM test (rA-t),
54.
S: TRA<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: aaaa<CR><LF>
The control character <ENQ> starts the test. It takes approximately one sec-
ond to complete the test.
6.3.38
TRS
Test RS232C interface. Test the RS232C interface. See Chapter 5.4.13
RS232C test (rS-t),
S: TRS<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
The control character <ENQ> starts the test. The test can be stopped by
pressing Ctrl-C.
Parameters
Significance
a
Torr lock
0 = Off (default)
1 = On
Parameters
Significance
aaaa
Error status. See Chapter 6.3.9 ERR,