4.1.2 Environment
Temperature:
Storage:
-20 ... +60 °C
Operation: +5 ... +50 °C (sea level)
+5 ... +30 °C (2000 m NN)
Relative humidity of the air:
max. 80 % (to 30 °C)
decreasing to max. 50 % (from 40 °C)
Usage :
indoors (max. 2000 m NN)
Contamination class:
II
Protection
class:
IP40
4.1.3 Standards
•
Compliance with Low Voltage Directive 2009/95/EG
•
Compliance with EMC Guideline 2004/108/EG
International/national standards as well as specifications:
•
EN 61010-1 (Electrical measuring and control equipment)
•
EN 61000-6-2 (EMC compatibility)
•
EN 61000-6-3 (EMC interference)
4.2 Mains
Connection
Voltage:
100…240 V AC
Frequency:
50/60
Hz
Device fuses:
2 x T1.6A H
Power consumption:
max. 5 VA
Overvoltage category:
II
Protection
class:
1
Connection:
Rubber connector IEC 320 C14
4.3 Measuring
Channel
Number:
1
Connection:
RJ45 (FCC 68)
Applicable sensors:
THERMOVAC
TTR90, TTR90S
TTR91, TTR91S
TTR96S
TTR211S
TTR216S
TTR100, TTR100S2
PENNINGVAC
PTR90
GA 09.034/2.02 – 07/10