Compact / Series 7600 Installation Procedure
12
Cover Removal
C-Arm
Rear Cover
1
.
Unplug the system power cord from the AC power receptacle.
2.
Remove the two screws located on the lower front section of the rear cover
and the screw located on each side of the cover.
3
.
Move the cover toward the floor. (This will release the top tab restraints.)
4
.
Pull the cover away from the system
.
5.
Remove the ground wire.
Workstation
Rear Cover
1.
Remove the ten screws securing the rear cover. Six along the sides and
four around the power cord plate.
2.
Pull the cover back away from the system.
3.
Remove the ground wire.
Mechanical Checks
In the steps below, check for ease of movement without excessive play.
1.
Check the Locking Pedals of the system for proper operation. Lock the
system in place by pressing the locking pedals down completely (Compact
/ Series).
2.
Check the Wig Wag movement of the system from side to side. Verify
proper operation of the wig wag lock.
3.
Check the Cross Arm movement of the system. Verify proper operation of
the cross arm lock.
4.
Check the Flip Flop movement of the system and integrity of the Pivot
Point. Check proper action of the pivot lock.
5.
Check the Radial Movement of the system and verify there is adequate
resistance to stop the C-Arm when the brake is applied.
6.
Check the Steering Handle of the C-Arm for proper operation.
Inspection and Setup
1. Verify or mount the skin spacer as directed by state regulation.
2. Inspect the AC power cable and interconnect cable for proper strain relief.
Inspect the power cable assembly and interconnect cable assembly for any
shipping damage. Look for any signs of damage or abrasion to the cable
jacket and inspect each plug for loose pins.