Compact / Series 7600 Installation Procedure
6
X-ray Source Assembly
Type
Fixed anode type
Focal Spot
Dual, 0.6 mm and 1.5 mm
Total Filtration
>3 mm minimum aluminum equivalent
Anode Heat
Storage Capacity
76,000 HU
Leakage
Technique Factors
110 kVp and 3.0 mA
Voltage Rating
110 kVp maximum
Fluoroscopy
Focal Spot Size
0.6 mm
High Voltage Range 36 kVp to 110 kVp, with manual adjustment
available.
Current Range
Standard 3 mA maximum
Dense Anatomy 6 mA maximum continuous
Snapshot 8 mA
Duty Cycle:
110 kVp @ 6 mA maximum continuous
Collimation
Nominal diameter iris circle for 9/6 inch system
6 inches (15.2 cm), 9 inches (22.8 cm)
Iris collimation is continuously adjustable to
less than 5 x 5 cm, measured at the SID.
Slot collimation is continuously variable and
rotatable.
Pulsed Fluorography Mode
Focal Spot
0.6 mm
kVp Range
36 kVp-110 kVp
mA Range
0.2 - 8.0 mA
Pulse Rate
1 pulse per second.