Compact / Series 7600 Installation Procedure
5
Input Power
NOTE:
The system operating voltage is changeable. If the operating voltage requires
changing, the change may only be accomplished by an OEC Medical Systems,
Inc. field service technician, or by qualified technical service personnel.
Line Frequency:
60/50 Hz
Single Phase:
110V ~
Standby (long-time) 5 A,
*Maximum (momentary) 25 A.
230V ~
Standby (long-time) 2 A,
*Maximum (momentary) 15 A.
*Maximum (Momentary) rating is for maximum radiographic exposure and
limited to 4 seconds.
Resistance of Mains 0.6 ohms/230 ~
0.3 ohms/110 ~
Technique Factors at Maximum Line Current Condition:
Peak current demand is independent of technique load factors.
Line Regulation (Percentage):
15% at maximum radiographic exposure. Also, under no conditions should the
loaded voltage fall below 100 VAC, measured on the secondary of the isolation
transformer.
Maximum Continuous Power Dissipation:
4,253 BTUs. Figure based on a maximum real power value of 1,245 watts.
NOTE:
Line regulation based on measurements made at the input (primary winding) of the
isolation transformer under standby conditions and at maximum radiographic
exposure by the following equation:
Line regulation = 100 (Vn - Vl)/Vl
Where Vn = no load voltage (standby)
Vl = loaded voltage (maximum radiographic exposure).