OEC 9800 Periodic Maintenance Procedure
4
General Instructions
Use this procedure in conjunction with the OEC 9800 Periodic Maintenance (PM) Inspection Report form. Enter the
required information at the top of the form prior to performing the PM test procedures. Read the “Safety” section.
Perform the procedures in the “Preparation” section and then begin the periodic maintenance. During testing, place a
check mark next to each test completed, and (where applicable) record the measured values; if adjustments to these
values are made, note them on the form in the comments section.
Note
:
The
numbered
sections and their titles in this document correspond to those on the form.
Log all test equipment used on the form in the block near the bottom,
Test Equipment Used
.
Note
:
Test equipment must have current calibration dates. Do not use test equipment with expired calibration
stickers.
Safety
Ingress Of Water And Solutions
The system must never be operated or stored in locations where conductive fluids, like water and saline solution,
might spill on the equipment.
Always unplug the AC power cable from the wall outlet before cleaning the equipment. Do not allow water, soap, or
other liquids to drip into the equipment and cause short circuits, electric shock and fire hazards.