OEC 9800 Periodic Maintenance Procedure
27
PS2
1.
Measure the voltages on the following power supply test points and record the values on the PM Form:
Note
:
The voltage designations on the Power Supply PCB are incorrect. Adjust the voltages to the value
described in the following table.
GND
to
Adjust
VDC
Tolerance
TP1
TP6
R7
-6.10
±
0.02
TP1
TP5
R31
+6.25
±
0.02
TP1
TP4
N/A
+14.0
±
1.0
TP1
TP3
R30
+24.1
±
0.02
TP1
TP2
N/A
-14.0
±
1.0