Fitting Guide

The VES

®

 is a monocular telescopic system. 

The choice of which eye to choose for the 
prescription of the VES

®

 in binocular patients 

is a function of which has the better visual 
acuity as well as the patient’s preferred 
sighting eye. The VES

®

 gives you the option 

to evaluate either eye, as it is completely 
interchangeable for the right or the left. 

The VES

®

 can be prescribed in the standard 

“above the frame” position or in the optional 
“low mounting” position. The choice of the 
appropriate positioning is a function of the 
intended use of the device as well as the 
anatomical features of the patient. Patients 
with wide or high bridges may prefer the low 
mounting position, as do users who intend to 
use it for intermediate and near rather than 
distance tasks. 

Bottom mounting is accomplished by 
attaching the telescope beneath the 
mounting plate rather than above it using 
the optional bottom mounting bracket as 
described in the four steps that follow:
(The standard mounting bracket will work
in many circumstances, however the 
optional bottom bracket will provide a wider 
range of adjustment.) 

1. Remove the VES

®

 from the

  bracket by unscrewing the allen  
  screws from each end. 

       

 

2. Remove the bracket from the   
  mounting plate by removing the
  two bottom screws. 

3. Reattach the bottom mounting  
  bracket beneath the mounting
  plate with the feet facing toward
  the back of the frame. 

       

 

4. Attach the telescope back onto
  the bracket with the back of the  
  telescope flush to the front of
  the bridge of the frame and the  
  bottom screws coming down
  through the top of the plate.
  Use of a screw-holding screwdriver
  is recommended.  

 

 

  (Available from Ocutech and    
  included in the diagnostic kit.)

The eyepiece should position

slightly above the pupil

shown here. 

Adjusting the focus of the VES

®

 is easy. 

The front focusing knob doubles as a 
focusing correction for the patient’s 
prescription as well as to focus the 
telescope for different distances. 

Demonstrate the convenient 
focusing of the VES

®

 

cautioning the patient not to 
block the front window opening 
of the telescope. 

Changing the Positioning of the Telescope
From Standard to Low Mounting Position

Focusing is Quick and Convenient 

Positioning the VES

®

 for the Proper Eye 

If the VES

®

 is not positioned

on the frame for the proper eye, 
remove the “side screws” with 
the allen wrench and flip the 
telescope over to position it for 
the proper side. Then re-insert 
the “side screws”. 

Frame should position 

slightly higher on the bridge.