14 Troubleshooting
74 / 86
Instruction Manual Mesotest II (G/62800/0715/en)
14 Troubleshooting
Attention
Risk of personal injury or equipment damage due to improper trouble-
shooting
If an error occurs which you are unable to correct by following the
instructions below, label the device as „out-of-order“ and contact
our service department or an authorized dealer.
Error
Possible source
Remedy
Nothing happens when the pow-
er switch is turned on
The Mesotest II is not connected
to the power supply
Power failure or power socket is
not active
The control pad or the serial ca-
ble of the computer is not con-
nected properly
The device fuses have blown
The unit has been switched off
and back on again too quickly
Plug the power cable into the
socket or into the low heat de-
vice socket at the Mesotest II.
Inform the in-house electrician
Check that the connector is
plugged in properly
)
Wait approx. 5 seconds before
turning the unit back on again
Patient claims he cannot see an-
ything
The unit is in stand-by mode
The examination room is too
bright
Press any button on the control
pad
Darken the room or cover the
back pane of the viewer with the
supplied cover.
Panel illumination is not func-
tioning
Lamp at the panel illumination is
faulty
R/L/BIN and -DPT. with
corresponding minus lens powers
are not working despite display in
the LCD
Supplementary equipment is not
installed in the device