19 Annex
Instruction Manual Easyfield® (GA/15000/XXXX/EN 1219 Rev01)
51 / 60
19 Annex
19.1 Electromagnetic Compatibility
Medical electrical equipment is subject to special precautionary
requirements with respect to EMC, and must be installed and operated
according to the EMC-Instructions contained in the accompanying
paperwork.
No special measures need be observed in respect of OCULUS devices and
systems.
Portable and mobile RF-communications devices can interfere with
electrically operated medical devices.
Minimal performance quality and essential performance criteria
A slightly disturbance of the analog camera of the Easyfield®
(slightly image noise on screen) during the examination is
permissible because it will not affect the diagnosis, treatment and
observation.
A short flicker of the illumination of the Easyfield® during the
examination is permissible because it will not affect the diagnosis,
treatment and observation.
A short interruption of the USB connection during the examination
is permissible because it will not affect the diagnosis, treatment and
observation.
Caution
The use of accessories, transducers and cables not specified by OCULUS
(for example as replacement parts) may result in increased emissions or
decreased immunity of the Easyfield®.
Use only the original accessories, transducers and cables specified by
OCULUS.
The use of accessories, transducers and cables specified by OCULUS with
devices other then the Easyfield® may result in increased emissions or
decreased immunity of the other device.
Do not use the accessories, transducers and cables specified by
OCULUS with devices other then the Easyfield®.