17 Declaration of Conformity
Instruction Manual Pentacam® / Pentacam HR® (G/70700/0109/en)
27 / 29
Страница 1: ...OCULUS PENTACAM PENTACAM HR INSTRUCTION MANUAL System for measuring and analysing the front part of the eye...
Страница 2: ...is basically identical This instruction manual describes how to manage patient data the default settings of the Pentacam program and the measuring process Additional functions pertaining to the Pentac...
Страница 3: ...e Pentacam off 8 7 Patient Data Management 8 7 1 Starting Patient Data Management 8 7 2 Entering new patients 9 7 3 Selecting existing patients 9 7 4 Starting the Pentacam program 10 8 Pentacam Progra...
Страница 4: ...R G 70700 0109 en 16 1 Terms of warranty 25 16 2 Assumption of liability for functions and damage 25 16 3 Address of the manufacturer and service department 26 17 Declaration of Conformity 27 18 Techn...
Страница 5: ...70900 07 001 Licence key dongle 70740 Testing log electrical safety Instruction Manual G 70700 HR G 70900 Dust protection cover 60100 5 1 Paper for chin support 65313 Software package and books Window...
Страница 6: ...o operating personnel at all times Observe the legal regulations with regard to accident prevention The Pentacam is a high quality technical product To ensure that it functions correctly and safely we...
Страница 7: ...acam in damp rooms Keep the Pentacam away from water that may drip surge or splash and make sure that no liquids can enter the Pentacam Do not place any containers with liquid ei ther close to or on t...
Страница 8: ...rate the Pentacam using original accessory parts supplied by us and when the device is in technically sound condition You must use the power supply belonging to the device in order to operate it There...
Страница 9: ...of 15 weeks 4 2 Disposal of used devices In accordance with Directive 2002 96 EC of the European Parliament and the Council of 27 January 2003 and in accordance with German law governing the circulat...
Страница 10: ...two seconds Up to 25 000 HR 138 000 genuine height values are measured and analysed At the same time any eye movement is recorded which the 3D model takes into account When the PC has received the ap...
Страница 11: ...operation Attention Before you use the device for the first time the OCULUS Pentacam examination area must be installed and connected by our service personnel or an authorised specialist Wait approx 3...
Страница 12: ...ng Patient Data Management After you have switched on the PC it first loads the operating system It then displays the user interface for Patient Data Management Fig 7 1 Patient Data Management To ente...
Страница 13: ...of the screen displays all previously examined patients in alphabetical order Click Search to quickly find the patient you require from the list Enter the patient s name or the first letter of the nam...
Страница 14: ...gram menu upper section Loading previous examinations Select the menu item Examination and click Load The dialog box Load Examination appears Make a selection by clicking the required examination Conf...
Страница 15: ...he eye currently being examined is detected automatically and is displayed in the Eye field 1 The Pupil Image 2 shows the vertical horizontal positioning of the Pentacam The object is to position the...
Страница 16: ...ective position of the camera and the eye which is currently being examined 9 1 1 Scheimpflug image settings You can set the number and type of images required for the respective examination in the Im...
Страница 17: ...while performing a full rotation twice around the eye Accommodation group box If you select this option the camera will take a total of 50 Scheimpflug images While the camera is recording the images...
Страница 18: ...al lenses gener al Enhanced dynamic for HR in the 3D scan examination mode 15 Yes If the pupil is insufficiently dilated apply mydriatic drops Use 3D scan for measurements Measuring functions 3D scan...
Страница 19: ...images 1 second Adjust the table height Ask the patient to place his or her head on the chin and forehead rest Rough adjustment The black ring marking between the chin rest and the forehead rest 1 sh...
Страница 20: ...oughly coincides with the circle R 3 on the sliding plate Fig 10 2 Markings on the cross slide Look at the patient s eye you are examining from one side and make sure that the blue slit light illumina...
Страница 21: ...l image right The image is sharpest when the red dot coincides with the red line in the Sche impflug image Focus the pupil image by moving the joystick towards the Pentacam or away from it Adjust the...
Страница 22: ...omati cally begin measuring alternately you can start the measuring procedure manually Fig 10 6 Adjustment window once the final position has been reached For measuring manually Start measuring by cli...
Страница 23: ...u have begun measuring either automatically or manually the Pentacam program opens The QS button appears in a field below the patient data Fig 11 1 Pentacam program with QS button If the QS button rea...
Страница 24: ...alysed Area If this value is less than the permissible threshold the patient must widen his or her eye Valid Data If this value is less than the permissible threshold you must darken the room Lost Seg...
Страница 25: ...t directory Transferring measurement data to another storage medium e g USB drive Sending stored data by e mail 12 1 Exporting patient data In Patient Data Management fig 7 1 page 8 select the patient...
Страница 26: ...f the program is updated on the Pentacam PC sender the program on the other PC recipient must also be updated 12 2 Importing patient data Import the patient and examination data you have received into...
Страница 27: ...please inform us or contact your authorized Oculus distributor to get further support 14 Troubleshooting Attention If an error occurs which you are unable to correct by following the instructions belo...
Страница 28: ...slit do not use a cloth or other cleaning agents Carefully clean the centre of the illuminated slit with purified compressed air Attention The optical components illuminating the slit and the lens in...
Страница 29: ...and have the damage confirmed on the consignment note to enable a proper claims settlement procedure Overall the general terms and conditions of business and delivery apply as per the date of purchase...
Страница 30: ...r authorised representatives will furnish you with additional information Address of the manufacturer and service department OCULUS Optikger te GmbH M nchholzh user Stra e 29 35582 Wetzlar Germany Tel...
Страница 31: ...17 Declaration of Conformity Instruction Manual Pentacam Pentacam HR G 70700 0109 en 27 29 17 Declaration of Conformity...
Страница 32: ...ee Speed 50 images in 2 seconds 500 measuring points HR 2760 per recorded image Number of evaluated measuring points Max 25 000 HR max 138 000 Dimensions H x W x D 535 x 280 x 360 mm Weight 9 kg Curva...
Страница 33: ...ock Protection class 1 Level of protection against electrical shock Type II B Level of protection against damaging water entry IP20 Temperature 10 C to 40 C Humidity 30 to 75 Air pressure 700hPa to 10...