distributor. 

Remarks:  If  use  the  detector  under  the  exceed  range,  this  is  not 

provided the free repair service. 

 

Calibrate service

 

We promise all detector in our company be calibrated with normal gas 

before leaving our factory, once bought from our company without need to 
calibrate, and must under the professional technical guidance. 

Warranty  is  12months  from  the  date  leaving  our  factory,  for  the 

accessory within 3months providing free maintenance, and for the detector 
providing the free calibration service once in a year. 

About the Backlight instruction

 

       

OC-903 with the electricity saver procedure, when the device without 

any  operation  within  30s,  the  detector  will  turn  on  the  electricity  saver 
procedure  automatically,  then  the  screen  will  become  dark,  and  press  any 
keys will become light again. 

Fast operation instruction

 

Mute: When the detector under the alarming, short press 

▲ 

or 

▼ 

will 

be muted.   

Manual storage

When the storage mode under the manual storage, on 

the  testing  interface  long  press  the 

,  then  after  the 

“beep”, will keep the measuring data. 

Menu  entrance

Under  the  testing  interface,  long  press  “ok”  for  5s 

then will enter the menu.

 

Parameter revise instruction

 

       

All  parameter  for  OC-903  could  be  revised  with  the  “Back”, 

“OK”,

▲,

▼.

 

Содержание OC-903

Страница 1: ...OC 903 Portable gas detector Operating Manual Henan Oceanus Import Export Co Ltd...

Страница 2: ...1 Catalogue Notice 2 Description 4 Feature 4 Parameter 5 Structure 7 Operation instruction 7 Operation interface 9 Remarks 12 Fault Treatment 13 Data logger function 14...

Страница 3: ...s OC 903 portable gas detector only available to detect the gas concentration when the pump on Treatment for the over measuring range Please avoid use the device in concentration which exceeds the mea...

Страница 4: ...in a year About the Backlight instruction OC 903 with the electricity saver procedure when the device without any operation within 30s the detector will turn on the electricity saver procedure automa...

Страница 5: ...er 32 bit microprocessor 24 bit ADC acquisition chip 2 4 inch IPS industrial grade color screen displaying each parameter It is able to select gas concentration unit of PPM VOL mg m3 For 1 4 gas detec...

Страница 6: ...gas type will be according to your requirements Temperature and humidity Temperature 40 120 Humidity 0 100 RH Testing mode Pump suction inner pump adjustable flow rage 100 1000mL L min workable for t...

Страница 7: ...et Gas List Some gas not list if need please contact manufacturer Gas Range Accuracy Resolution Response time Ex 0 100 LEL 3 F S 0 1 LEL 20s CH4 0 100 LEL 3 F S 0 1 LEL 20s O2 0 30 VOL 3 F S 0 01 VOL...

Страница 8: ...esting interface Menu interface Setting interface Six keys will have different function in the three interfaces as the following Testing interface Menu interface Setting interface move to left or up m...

Страница 9: ...ion when the pump on 5 2 Power on When power off long press key for 5s to activate the backlight and red indicator then beep then the detector power on Then the display will be sensor testing 3s senso...

Страница 10: ...nt interface as the following picture displaying one gas two gas three gas four gas orderly As the above displaying time temperature pump running battery and humidity Take A for example for O3 CO CH2O...

Страница 11: ...can turn on or off some one or more channel gas With the Pump flow rate can adjust the pump flow rate only need to setting under vacuum environment default is 30 With Temperature correction and Humidi...

Страница 12: ...r into the target gas calibration interface slowly turn on the calibrate gas and control the flow rate within 500ml min observing the detector gas concentration now should be rising when the value up...

Страница 13: ...e and round cycle time between the twice timing detecting Setting the timing detection need to back to the Detecting interface to press the Run stop key turn on the countdown to begin timing detection...

Страница 14: ...e to normal value If still not normal then zero calibrate in clean air Response weak Reason Lower level of oxygen 5 VOL High pressure pump not work The sensor useful life end or other faults Solution...

Страница 15: ...if not then need to change the charger if is 5V then need to send back to manufacturer for testing 9 Data logger OC 903 with the data logger function the charging port is the data output port the oper...

Страница 16: ...time of history data should be in this period range 6 Click the search can find the above period gas concentration for the according channel gas the data also could be saved as the EXCEL form by click...

Отзывы: