3: Using NeoFox with NeoFox Viewer Software
013-20000-009-02-201510
25
8. Click on
button on the bottom right corner of the screen. The
Computed
Calibration Constants
table fills in on the bottom left corner of the screen.
9. Click the
button to use the coefficients for oxygen calculation. A Download
Confirmation screen appears.
10. Click Yes to confirm the download and continue.
Two-Point Calibration
If you want to do your own calibration instead of using a factory calibration, follow these Two-Point
Calibration instructions.
►
Procedure
1. From the Main screen, select
Options | Calibration
.
2. Click the
Setup
tab. The Setup screen appears.