
7: Calibration
225-00000-000-11-201604
63
14. Click on the
Save All Values
button to save the information, and then
Exit
the USB
Programmer software.
The new wavelength calibration coefficients are now loaded onto the EEPROM memory chip on
the Flame.
Irradiance Calibrations
Irradiance calibrations and relative irradiance calibrations are about quantifying the spectra, by
translating the signal (incident number of photons) to a calibration. This can be either absolute
(an atomic emission light source of known output power) or relative (corrected for instrument
response function but not absolute units). It can be considered a measurement technique and
is used widely in remote sensing, light metrology and anywhere where you wish to characterize
the incident light source. Irradiance calibrations are not required for many techniques because
these measure the relative signal changes with respect to the sample and not the light source.
You can find out more about irradiance calibration techniques at
http://oceanoptics.com/measurementtechnique/irradiance/
OceanView has wizards that will step you through absolute irradiance and relative calibrations
and more information on these is located in the OceanView manual (see
).
Содержание Flame-S
Страница 18: ...1 Introduction 8 225 00000 000 11 201604...
Страница 32: ...2 Installation and Setup 22 225 00000 000 11 201604...
Страница 46: ...4 Troubleshooting 36 225 00000 000 11 201604...
Страница 54: ...5 How the Flame Spectrometer Works 44 225 00000 000 11 201604...
Страница 58: ...6 Technical Specifications 48 225 00000 000 11 201604 Mechanical Diagram Figure 1 Flame Outer Dimensions...
Страница 74: ...7 Calibration 64 225 00000 000 11 201604...
Страница 130: ...Index 120 225 00000 000 11 201604...