Oakton Temp-16 Скачать руководство пользователя страница 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Instruction Manual 

 

Temp-16  

Precision RTD Thermometer 

68X481904 Rev 0  09/09

Part of Thermo Fisher Scientific 

Содержание Temp-16

Страница 1: ...Instruction Manual Temp 16 Precision RTD Thermometer 68X481904 Rev 0 09 09 Part of Thermo Fisher Scientific...

Страница 2: ......

Страница 3: ...ns 10 9 Display Overview 11 10 Measurement Mode 12 11 Selecting Temperature Scale 13 12 Min And Max Function 14 13 Auto Off Function 14 14 Calibration 15 15 Clear Field Calibration 16 16 Field Calibra...

Страница 4: ......

Страница 5: ...n backlight 3 Select C or F display using C F key 4 Insert probe in sample Allow adequate time for reading to stabilize Response time will vary depending on probe 5 Take reading Press hold key to free...

Страница 6: ...peration which includes the following features Operator selection of Celsius or Fahrenheit scale Resolution of 0 1 auto ranging Four digit backlit LCD Hold feature to retain a reading temporarily Fiel...

Страница 7: ...is instrument is designed to accept low level signals supplied by standard 100 RTDs Under no circumstances should the input voltage exceed the specified 10 v rms 1 Do not use or store this instrument...

Страница 8: ...o 1562 F Display 4 digit LCD X digits 2 1 3 X 1 backlit viewing area Display Update Rate 0 5 sec per update Input One probe with 3 pin circular connector Switchcraft TA3F Input Protection 10V rms Batt...

Страница 9: ...5 Weight Without armour With armour 267 g 362 g Ingress Protection With Probe Attached Meets IEC 529 IP 54 for dust and water resistant enclosures CE Compliance EN61326 1 A1 1998 EU EMC Directive...

Страница 10: ...nd cause damage to unit The typical battery life is about 700 hours Selected settings are stored in memory and will remain in memory even after power is turned off or while batteries are being replace...

Страница 11: ...o the optional rubber armor slide in from the top of meter before pushing the bottom edges of meter down to set it into position Lift up the stand at the back of meter for bench top applications if ne...

Страница 12: ...8 6 ASSEMBLING OPTIONAL HANDSFREE ACCESORIES You can use the optional magnets and strap in the Hands Free Kit accessories for hands free operations Figure 2 Hands Free Kit...

Страница 13: ...orrect 100 RTD alpha 0 003850 for your instrument Using an incorrect probe type will result in erroneous readings Insert the 3 pin plug into the mating connector on the top of the instrument Figure 3...

Страница 14: ...ing of the measured reading while in measurement mode min max Toggles between minimum and maximum readings MIN and MAX readings are calculated from the last min max keypress Press and hold this key fo...

Страница 15: ...annunciator MAX Maximum reading annunciator 3 Bar 700 550 hours 2 Bar 350 550 hours 1 Bar 150 350 hours Blinking 150 hours CAL Blinks during calibration mode Remains in display indicating field calibr...

Страница 16: ...mum instrument accuracy allow one minute for ambient temperature stabilization If the unit has been stored at an extreme ambient condition more time will be needed A STABLE indicator appears at the to...

Страница 17: ...Select C or F by pressing the C F key Each time the key is pressed the temperature scale will switch Switching between C and F can be done at any time during operation Each time you turn the instrumen...

Страница 18: ...maximum values are sensed and automatically stored To exit and clear the MIN MAX reading stored press and hold the min max key for 3 seconds 13 AUTO OFF FUNCTION The thermometer has a default auto off...

Страница 19: ...brate 1 Pack sensing end of probe in a container tightly packed with crushed ice and filled with distilled water Allow temperature to stabilize 2 Press and release the cal key for 3 seconds to enter t...

Страница 20: ...essfully cleared 16 FIELD CALIBRATION UNLOCK The calibration unlock feature enables field calibration operation To unlock field calibration 1 Turn the thermometer off 2 Hold the cal key and hold key d...

Страница 21: ...he lockout remains in effect until an unlock has been performed To lock the field calibration operation 1 Turn the thermometer off 2 Hold the cal key and C F key down simultaneously then press the on...

Страница 22: ...definitely and not require service other than occasional cleaning of the housing and changing of the batteries Do not clean with abrasives or solvents Use mild detergents never immerse nor use excessi...

Страница 23: ...ct polarity OPEN display on LCD Probe not connected Make sure probe is firmly connected Ur or Or display on LCD Measurement over Or or Under Ur range Ensure temperature taken is within meter s specifi...

Страница 24: ...85 General purpose probe 10 in long x 0 188 in dia 10 second time constant 08117 70 PTFE coated general purpose probe 10 in long x 0 145 in dia 15 second time constant 08117 87 Penetration probe 4 in...

Страница 25: ...outside the warranty period or those resulting from product misuse or abuse may be invoiced to you 24 PRODUCT RETURN To limit charges and delays contact the seller or Manufacturer for authorization a...

Страница 26: ...formity ISO IEC 17025 2005 accredited NIST Handbook 150 2000 Edition ANSI NCSL Z540 2 1997 NIST Technical Note 1297 ISO 9000 2000 Fast Service Our substantial inventory of replacement parts ensures a...

Страница 27: ...identification of the item certified condition of the item issue date identification of calibration procedure calibration date as found as left test data where applicable signature of technician state...

Страница 28: ...tios at no extra cost Call us today and see why InnoCal is The Choice of Quality See our Scope of Accreditation for any limitations Calibration test points against NIST traceable standards Meter only...

Страница 29: ...ized seller For more information on OAKTON Instruments Products please contact your nearest distributor or visit our web site listed below OAKTON Instruments P O Box 5136 Vernon Hills IL 60061 USA Tel...

Страница 30: ......

Страница 31: ......

Страница 32: ...Distributed by...