5) Place the 20.0 NTU calibration vial in the sample 

  vial holder, press                  to start 20 NTU calibration. 

After calibration is done, press                    to confirm. The   

instrument will return to the calibration menu (see 

diagram on the right side). The OK sign indicates that   

20 NTU has been calibrated 

 

6) If you want to continue to calibrate 100, 400 and 800 NTU, place the 100 NTU calibration vial in 

the  sample  vial  holder  and  press                to  calibrate.  After  calibration  is  done,  press                to 

confirm; repeat the steps to calibrate 400 NTU and 800 NTU. After calibration is done, press                                                       

to return to measurement mode. 

7)  Press            to  exit  calibration  mode,  the  instrument 

will  return  to  measurement  mode  as  shown  in  Diagram 

on the right side.

 

 

 

4.3  Notes for Calibration 

a) 

Calibration point verification:

 The calibration point can be verified after the calibration is 

completed. If the calibration point has a large error, enter the calibration mode and repeat the 

calibration. For calibration point accuracy, users can refer to the following standards: 

 

 

 

 

 

 

b) 

Calibration point selection: 

The instrument has been calibrated at full range before leaving the 

factory. For subsequent use, you can select 2 or more points as needed, as long as the estimated 

measurement range is between the two calibration points. In calibration setup mode, press          or             

         

to select the calibration point. 

c) 

Low turbidity calibration requirement

 

 

For low turbidity measurements (measurement less than 2 NTU), please test 0.0 NTU 

calibration solution. If the accuracy is not meeting requirements, calibrate the instrument at 0.0 

NTU and 20.0 NTU before test; then use 1

#

 or 2

#

 vial for measurement.   

Calibration point 

Accuracy for reference 

0 NTU 

≤0.05   NTU 

20 NTU 

≤±0.2   NTU 

100 NTU 

≤±2 NTU 

400 NTU and 800 NTU 

≤±5 NTU 

Содержание T100WL

Страница 1: ...1 T100WL Portable Turbidity Meter 35635 10 Instruction Manual...

Страница 2: ...eparation for Calibration 10 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 11 4 3 Notes for Calibration 12 5 Turbidity Measurement 13 5 1 Sample Vial Handling 13 5 2 Measurement Prepar...

Страница 3: ...parameter setting TruRead measurement mode automatically takes multiple consecutive readings and calculate their average minimum and maximum value as well as displaying each set of data in a brief te...

Страница 4: ...rmazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measuring Mode Normal measurement and TruRead measurement Data Storage 200 groups Data Output USB to PC Calibration Record Ca...

Страница 5: ...er when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark with the mark on the cali...

Страница 6: ...ofiber cloth Power adaptor 5V 1A Software flash drive Silicone oil 10 ml Sample vials 6 USB Cable under the meter User Manual Diagram 2 Measurement Mode Measurements and unit Operation Guide Battery i...

Страница 7: ...iagram 3 Calibration Mode Standard values Operation guide Finished calibration indicator Parameter Setup Menu Operation guide Data Recall Saved data numberings Saved data Operation guide Calibration M...

Страница 8: ...nt while it is being charged If charging the instrument when it is turned off Charging will be displayed and Charging is completed will be displayed after a full charge 3 6 Data Log Recall and Deletio...

Страница 9: ...surement mode Measurement Mode Storage Capacity Normal 200 sets TruRead 5 times measurement 100 sets TruRead 10 times measurement 61 sets TruRead 15 times measurement 44 sets TruRead 20 times measurem...

Страница 10: ...0 NTU calibration solution and close the vial cap Replace laboratory distilled water open the vial cap pour out the original solution add 1 2 distilled water cover the cap and shake the vial to rinse...

Страница 11: ...ake the solution even But sediment can still occur during tests and make the measurement unstable Users need to have rich experience with Formazin calibration solutions 4 2 Calibration Procedure Take...

Страница 12: ...n The calibration point can be verified after the calibration is completed If the calibration point has a large error enter the calibration mode and repeat the calibration For calibration point accura...

Страница 13: ...e replace the 0 0 NTU solution in time according to the requirements in Section 4 1b e High turbidity calibration requirement For turbidity measurement greater than 2 NTU it is recommended to calibrat...

Страница 14: ...number of continuous measurements is set to 20 times in parameter setting P2 a measured value will be displayed every 20 seconds Diagram 9 2 is the third measurement Diagram 9 3 is the display interf...

Страница 15: ...c for details c Mixing and Degassing Samples should not be vigorously shaken or vibrated It is recommended that users gently shake the sample vial to make solution evenly distributed Air bubbles in s...

Страница 16: ...lect in how soon the instrument will turn off automatically if there is no operation 10 minutes 20 minutes 30 minutes or off when selecting off the instrument can only be turned off by pressing the po...

Страница 17: ...er setting information Function buttons 7 2 Operation Key in Software Interface Refresh Port reset key When the instrument and computer are not connected click to connect SyncTime Sync time key click...

Страница 18: ...ed in the USB drive 7 4 Run Software Click Download key all data in the instrument memory will be sent to the computer When the program is running all measurement information will be sent to the compu...

Страница 19: ...are as in Diagram 13 Diagram 14 shows you how to replace the lamp Unscrew the lamp cover with a coin Use a tweezer to unplug the socket Use a tweezer to unscrew the lamp cover Take out the old tungste...

Страница 20: ...limited warranty Within the warranty period repair parts or product replacement will be made without charge due to malfunctions caused by manufacturing defects 9 2 The above limited warranty shall not...

Отзывы: