Oakton T100WL Скачать руководство пользователя страница 10

 

- 10 - 

3.7  Setting for first-time use. 

The instrument should be fully charged before first-time use. Check and adjust settings for the following 

items: date, time, system language, etc. See the detail in Section 6 Parameter setup. 

CALIBRAT ION

 

4.1  Preparation for Calibration 

a)  Calibration point 

The instrument has 5 calibration points: 0 NTU, 20 NTU, 100 NTU, 400 NTU, and 800 NTU. Among 

them, 0 NTU point uses AMCO 0.0 NTU calibration solution or laboratory distilled water, and the 

remaining 4 calibration points use AMCO polymer solutions. Note that the cap of the 0.0 NTU 

solution vial can be unscrewed. After the solution is invalidated, users can replace the 0.0 NTU 

calibration solution or laboratory distilled water. The 0.0 NTU calibration solution can be purchased 

from the supplier; For the remaining 4 calibration solutions, their vial caps cannot be opened. Simply 

dispose the solutions after they are expired and buy new ones from your supplier to replace. 

b)  Replace zero turbidity solution 

 

Replace 0.0 NTU calibration solution: open the vial cap, pour out the original solution, add 

1/2 distilled water, cover the cap and shake the vial to rinse it and pour out the water. 

Repeat it 3 times. Shake off the distilled water in the vial. Pour in new 0.0 NTU calibration 

solution and close the vial cap. 

 

Replace laboratory distilled water: open the vial cap, pour out the original solution, add 1/2 

distilled water, cover the cap and shake the vial to rinse it and pour out the water. Repeat it 3 

times. Shake off the distilled water in the vial. Pour in new laboratory distilled water and 

close the vial cap.   

 

The accuracy of 0.0 NTU calibration solution and laboratory distilled water is same. But 0.0 

NTU calibration solution has 6 to 12 months of shelf-life, distilled water can only be used for 

several days. 

c)  Clean vial surface 

Apply a small drop of silicone oil on the surface of the vial and wipe it off with a micro-fiber cloth to 

evenly distribute the silicone oil on the surface so as to cover smudges or scratches, which helps 

light scattering. But please pay special attention to the following points: 

  The silicone oil applied should not be too much. After wiping with micro-fiber cloth, please wipe 

with filter paper or high-quality tissue paper to clean off. Excessive residual silicon oil on the vial 

surface will affect the measurement accuracy. 

  It's not necessary to use silicone oil for each calibration and measurement. Apply silicone oil 

every several days or once a week. In between, just clean the surface with filter paper or 

high-quality tissue paper. 

  Clean the calibration vial and sample vial together and keep the steps and actions consistent to 

achieve same degree of cleanliness. 

Содержание T100WL

Страница 1: ...1 T100WL Portable Turbidity Meter 35635 10 Instruction Manual...

Страница 2: ...eparation for Calibration 10 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 11 4 3 Notes for Calibration 12 5 Turbidity Measurement 13 5 1 Sample Vial Handling 13 5 2 Measurement Prepar...

Страница 3: ...parameter setting TruRead measurement mode automatically takes multiple consecutive readings and calculate their average minimum and maximum value as well as displaying each set of data in a brief te...

Страница 4: ...rmazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measuring Mode Normal measurement and TruRead measurement Data Storage 200 groups Data Output USB to PC Calibration Record Ca...

Страница 5: ...er when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark with the mark on the cali...

Страница 6: ...ofiber cloth Power adaptor 5V 1A Software flash drive Silicone oil 10 ml Sample vials 6 USB Cable under the meter User Manual Diagram 2 Measurement Mode Measurements and unit Operation Guide Battery i...

Страница 7: ...iagram 3 Calibration Mode Standard values Operation guide Finished calibration indicator Parameter Setup Menu Operation guide Data Recall Saved data numberings Saved data Operation guide Calibration M...

Страница 8: ...nt while it is being charged If charging the instrument when it is turned off Charging will be displayed and Charging is completed will be displayed after a full charge 3 6 Data Log Recall and Deletio...

Страница 9: ...surement mode Measurement Mode Storage Capacity Normal 200 sets TruRead 5 times measurement 100 sets TruRead 10 times measurement 61 sets TruRead 15 times measurement 44 sets TruRead 20 times measurem...

Страница 10: ...0 NTU calibration solution and close the vial cap Replace laboratory distilled water open the vial cap pour out the original solution add 1 2 distilled water cover the cap and shake the vial to rinse...

Страница 11: ...ake the solution even But sediment can still occur during tests and make the measurement unstable Users need to have rich experience with Formazin calibration solutions 4 2 Calibration Procedure Take...

Страница 12: ...n The calibration point can be verified after the calibration is completed If the calibration point has a large error enter the calibration mode and repeat the calibration For calibration point accura...

Страница 13: ...e replace the 0 0 NTU solution in time according to the requirements in Section 4 1b e High turbidity calibration requirement For turbidity measurement greater than 2 NTU it is recommended to calibrat...

Страница 14: ...number of continuous measurements is set to 20 times in parameter setting P2 a measured value will be displayed every 20 seconds Diagram 9 2 is the third measurement Diagram 9 3 is the display interf...

Страница 15: ...c for details c Mixing and Degassing Samples should not be vigorously shaken or vibrated It is recommended that users gently shake the sample vial to make solution evenly distributed Air bubbles in s...

Страница 16: ...lect in how soon the instrument will turn off automatically if there is no operation 10 minutes 20 minutes 30 minutes or off when selecting off the instrument can only be turned off by pressing the po...

Страница 17: ...er setting information Function buttons 7 2 Operation Key in Software Interface Refresh Port reset key When the instrument and computer are not connected click to connect SyncTime Sync time key click...

Страница 18: ...ed in the USB drive 7 4 Run Software Click Download key all data in the instrument memory will be sent to the computer When the program is running all measurement information will be sent to the compu...

Страница 19: ...are as in Diagram 13 Diagram 14 shows you how to replace the lamp Unscrew the lamp cover with a coin Use a tweezer to unplug the socket Use a tweezer to unscrew the lamp cover Take out the old tungste...

Страница 20: ...limited warranty Within the warranty period repair parts or product replacement will be made without charge due to malfunctions caused by manufacturing defects 9 2 The above limited warranty shall not...

Отзывы: