Operating Manual

pHTestr

®

 BNC 10 Testers

Keypad Functions
Short press 

= <2 seconds

   Long press 

= >2 seconds

1. When meter is turned off, short press to turn on the 

    meter and long press to enter parameter setting mode.

2. In measurement mode, long press to turn off the 

    meter and short press to turn on/off the back light.

3. In manual temperature adjustment mode, short press to 

    scroll up the digit one by one and long press to scroll 

    rapidly.

4. In parameter setting mode, short press to scroll up the 

    digit one by one and long press to scroll rapidly  (Only in 

    parameter setting P2 and P3).  

1. In measurement mode, short press to switch 

    parameter from pH to mV. Long press to enter manual 

    temperature adjustment.

2. In manual temperature adjustment mode, short press 

    to return to measurement mode.

3. In calibration mode, short press to confirm calibration 

    and return to measurement mode.

4. In parameter setting mode, short press to confirm 

    parameter and long press to enter measurement mode.

1. In measurement mode, long press to enter calibration 

    mode.

2. In manual temperature adjustment mode, short press 

    to scroll down the digit one by one and long press to 

    scroll raplidly. 

3. In calibration mode, short press to cancel calibration 

    and return to measurement mode.

4. In parameter setting mode, short press to scroll down

    the digit one by one and long press to scroll rapidly 

    (Only in parameter setting P2 and P3).

5. When the measured value is locked (HOLD icon), short 

    press to unlock.

 

Thank you for purchasing an Oakton

®

 pHTestr

®

 Pocket pH Tester.

Please carefully read this operating manual before using the 

product to obtain an accurate and reliable test result and avoid 

unnecessary damage to the tester or sensor/probe.

Calibration

1-point

2-point

3-point

USA series

7.00 pH

7.00 pH,

4.00 or 

1.68 pH,

10.01 or

12.45 pH

NIST series

6.86 pH

6.86 pH,

4.01 or

1.68 pH, 

9.18 or

12.45 pH

Indication

icons 

L M H

Recommended

Accuracy:≥0.1 pH

7.00 pH,

4.00 pH or

1.68 pH

6.86 pH,

4.01 pH or

1.68 pH

L M

Range:<7.00 pH

7.00 pH,

10.01 pH or

12.45 pH

6.86 pH,

9.18 pH or

12.45 pH

M H

Range:<7.00 pH

Range:

0 to 14.00 pH

Symbol Self-diagnostic 

information

Checking and how to fix

1. Check whether pH buffer solution

    is correct (1st point calibration 

    must be 7.00).

2. Check whether the probe is 

    damaged.

3. Check if there is any air bubble in

    the glass bulb sensor.

Wrong pH buffer 

solution or the range 

of calibrationsolution 

exceeds standard.

Press CAL/

  key 

when reading is not 

stable during

measurement.

Wait for the measurement

stability icon (

) to appear and

stay, then press CAL/

  key.

pH Calibration

1. Short press    /

 key to turn on.

2. Connect the pH electrode with the BNC connector, rinse the 

probe in distilled water and use tissue paper to gentlydab off 

excess water.

3. Pour certain amount (about half volume of the calibration bottles) 

of pH 7.00 pH and pH 4.00 buffer solutions in the corresponding 

calibration bottles.

4. Long press CAL/

 key to enter calibration mode, short press CAL/

 

key to cancel calibration and return to measurement mode.

5. Dip the probe in pH 7.00 buffer solution, 

stir gently, and allow it to stand still in the 

buffer solution until a stable reading is 

reached. When measurement stability 

icon (

) appears and remains on the 

screen (see Fig 1); then short press 

MODE/

  key to complete the one-point 

calibration. The tester returns to 

measurement mode and indication icon 

M

” will appear at the bottom left of the 

screen, indicating the middle point of 

calibration has been completed.

6. For 2-point calibration, start with pH 7.00 buffer solution; once 

confirmed then rinse the probe in distilled water and dry it with 

tissue paper. Dip the probe into pH 4.00 buffer solution, follow the 

steps above to complete 2-point calibration. Indication icons “

L

” 

and “

M

” will appear on the bottom left of the screen.

7. If necessary to perform a 3-point calibration, rinse probe with 

water, dip the probe into pH 10.01 buffer solution (sold 

separately), and follow the steps above. Indication icons “

L

”, “

M

”, 

and “

H

” will appear on the bottom left of the screen.

Notes

A. Tester can perform 1 to 3 points automatic calibration. Please 

note that pH 7.00 (USA Standard) or pH 6.86 buffer solution 

(NIST Standard) must be used to conduct 1st point calibration. 

Then use other buffer solution to conduct 2nd or 3rd point 

calibration. Tester will automatically recognize five kinds of pH 

buffer solutions. For details, please refer to the following table:

B. The tester has self-diagnostic functions:

* The 1st point calibration must be 7.00 pH. Perform the 2nd point 

calibration (4.00 pH) immediately after the 1st point. Do NOT turn 

off the meter before you conduct 2nd point calibration. If the meter 

is turned off after 1st point calibration, user will need to restart the 

calibration process with the 7.00 pH first and the 4.00 pH following 

after. Calibrating directly in pH 4.00 after turning meter off and 

back on will cause “Er1”.

pH Measurement

1. Short press    /MEAS key to turn on the tester. 
2. Connect the pH electrode with the tester. Rinse the probe in 

distilled water, and gently dab off excess water with tissue paper 

(do not rub or wipe the glass sensor).

3. Dip the probe in sample solution, stir gently, and allow it to stand 

until a stable reading is reached. Take readings after stability 

indicator icon (

) appears and stays on the screen.

Special Notes

• The pH probe must be rinsed thoroughly after each use. Soap 

water should be used to clean off any grease or other 

contaminants.

 

Refer to your pH electrode manual for important information 

regarding electrode maintenance. 

 

• Do NOT store probe in distilled or deionized purified water 

because that will cause permanent damage to the pH probe. 

Purified water is only recommended for rinsing the probe. The 

probe should be stored in 3M KCL pH electrode storage solution 

(SKU 00653-04) for best accuracy.

Features 

• Large easy-to-read LCD with 3-color backlight

• Accuracy of ±0.01 pH ± 1 digit

• 1, 2, or 3 calibration points

• Auto buffer recognition: US or NIST

• Stability and battery status icons

• Auto power-off function conserves battery life

• IP67 waterproof rating

Setting the Parameters

When tester is turned off, long press    /

△ 

key to enter parameter 

setting. Short press    /

△ 

or CAL/

▽ 

key to switch from P1 to 

P2…P7. Short press MODE/

   key and parameter will flash, then 

short press     /

△  

or CAL/

▽ 

key to choose desired parameter, 

press MODE/

   key to confirm. Long press MODE/

   to confirm 

and return to measurement mode. In P2 or P3, short press    /

 or 

CAL/

▽ 

to change the digit one by one and long press to change 

rapidly.