Operating Instructions

  

ORPTestr

®

 50 Pocket Tester

Temp Set
Backlight
Reset

hold

Backlight 

ON

 

OFF

hold

Backlight

OFF

hold

menu

Backlight 

Reset

Temp Cal

hold

°C

 

°F

 

menu

 °C

hold

Temp Set

Set °C/°F

hold

Getting Started

The ORPTestr 50 Pocket Tester has been factory calibrated and usually works 

well out of the box. However, after extended periods of non-use, it is best to 

remove the sensor cap and soak the sensor in warm tap water or ORP buffer 

for 10 minutes or so. A brief rinse with deionized (DI) water is OK, but avoid 

soaking or storing in deionized water as this will shorten ORP electrode life. 

Prior to taking the measurements, periodic calibration with certified standards 

is recommended for best accuracy.

Temperature Settings

1. Press ON/OFF ( ) to power on if needed.

2.  Press MENU/

v

 

to enter setup window. Press HOLD/

  

to select Temp Set.  

The display shows Set °C/°F  and Temp Cal.

3.   Press  HOLD/

  

to select Set °C/°F. Display automatically shows °C and °F.

4. Scroll down by pressing MENU/

v

 

to toggle between °C and °F.

5.  Press HOLD/

  

to select a temperature unit. The display shows the selected  

temperature setting with a  . 

Temp Set

Settings

Factory Reset

hold

No

 

Yes

menu

hold

User Reset

Reset

hold

• 

Aquaculture 

• Aquariums

 

• Boilers

 

• Car washes 
• Ecology 

 

Applications

• Food sectors 

 

• Hydroponics

 

• Labs

 

• Sanitation plants

 

• Steam generators

 

• Spas 

 

• Studies

 

•  Water and wastewater 

treatment

•  Water quality testing

 

in pools and more 

Backlight Settings

1. Press ON/OFF ( ) to power on if needed.

2.  Press MENU/

v

 

to enter setup window. Scroll down by pressing 

MENU/

v

 

to select Backlight. Press HOLD/

  

to select Settings. The 

display shows Backlight.

3. Press HOLD/

  

to select Backlight. The display shows ON and OFF.

4.  Scroll down by pressing MENU/

v

 

to toggle between ON and OFF. 

Backlight ON increases readability in low-light conditions.

5.  Press HOLD/

  

to select the desired backlight option. The display 

shows the selected backlight option with a  .

 

CAL

  
   

189 

    189

 

mV

 

CAL

Place the tester in 
the buffer to  
start or press                 
to escape.

cal

esc

cal

esc

 

CAL

  
   

844 

   25.0°C

 

 

RmV

ORP Measurement

1. Press ON/OFF ( ) to power on if needed.

2.  Dip the electrode about 2 cm to 3 cm into the test solution. Stir and let the 

reading stabilize. The timer icon blinks during this time. Once the reading is 

stabilized, the timer stops blinking and   will appear to indicate the stability 

of the reading. 

 

CAUTION: Testing dry samples is not accurate and can  

lead to sensor damage or breakage. Soils must be wet  

and free of particulates that may scratch the glass  

sensor. Excessive force into dry samples can cause  

glass breakage.

3.  Note the ORP value or press HOLD/

  

to freeze the reading. To release the 

reading, press HOLD/

  

again.

4.  Press ON/OFF ( ) for 3 seconds to turn off tester. If a key is not pressed for  

8.5 minutes, the tester will automatically shut off to conserve batteries.

 

ORP

  

   

844

 

   

25.0°C

RmV

ORP Calibration

Calibration should be done regularly, preferred once in a week. ORP tester 

allows one point calibration.

1. Press ON/OFF ( ) to power on if needed.

2.  Dip electrode in about 2 cm to 3 cm into the calibration standard solution.

3.  Press the CAL/ESC key to enter calibration mode. The CAL indicator will be  

displayed.

4.  Allow the reading to stabilize. The timer icon blinks during the reading  

stabilization. Once the reading stabilized the timer stops blinking.

5.  Upper display shows current measured value based on the factory default.

6.  Lower display shows current measured mV with last calibration (if no  

calibration is performed earlier both display shows the same measurement 

value).

7.   Press MENU/

v

 

to adjust the mV value to match the calibration standard 

value.

8.   Press HOLD/

  

to confirm the calibration or press CAL/ESC key to exit the  

calibration without confirmation.

User Reset

Reset the ORP calibration to the default using the user reset function.

1.  Press MENU/

v

 

to enter setup window. Scroll down by pressing MENU/

v

  

to select Reset. Press HOLD/

  

to select Reset. The display shows User 

Reset and Factory Reset.

2.  Press HOLD/

  

to select User Reset. The display automatically shows  

No

 

and

 

Yes.

3. Scroll down by pressing MENU/

v

 

to toggle between No

 

and

 

Yes

.

4.  Press HOLD/

  

to confirm either No or Yes. The display shows the user  

reset option with a  .

Factory Reset

Reset the tester to the factory default using the factory reset function.

1.  Press MENU/

v

 

to enter setup window. Scroll down by pressing the 

MENU/

v

 

to select Reset. Press HOLD/

  

to select Reset. The display shows  

User Reset and Factory Reset.

2.  Scroll down by pressing the MENU/

v

 

to toggle between the resets.  

Press HOLD/

  

to select Factory Reset. The display automatically shows  

No and Yes.

3. Scroll down by pressing MENU/

v

 

to toggle between No and Yes.

4.  Press HOLD/

  

to confirm either No or Yes. The display shows the factory 

reset option with a  .

Temp Set
Settings 

User Reset

hold

No

 

Yes

menu

hold

Reset

Factory Reset

hold

Manual Temperature Setting

The MTS range is 0°C to 60.0°C (32.0°F to 140.0°F). User reset will set  

temperature to default value 25°C or 77°F.

1.  Press MENU/

v

 

to select setup window. Scroll down by pressing MENU/

v

 

to select Temp Set. Press HOLD/

  

to select Temp Set. The display shows  

Set °C/°F and Temp Cal.

2.  Scroll down by pressing MENU/

v

 

to toggle between Set °C/°F and  

Temp Cal. Press HOLD/

  

to select Temp Cal.

3.  The lower display shows the current measured temperature reading based 

on the last set offset and the upper display shows the current measured 

temperature reading based on factory default calibration.

4.  Dip the tester into a solution of known temperature and allow time for the  

built-in temperature sensor to stabilize.

5.  Press MENU/

v

 

to adjust the temperature value or press the HOLD/

  

to  

confirm the calibrated value as new temperature value of the solution.

Note: To exit without confirming the calibration, press CAL/ESC.

 

Backlight 

Reset

hold

Temp Cal 
23.4°C 
23.4°C

menu

23.4°C

hold

Temp Set

Set °C/°F

Temp Cal

hold

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

Key Info   

Scroll 

Accept 

EXIT    

  

MENU 

HOLD 

CAL   

1065O100_MAN_35634-45.indd   1

9/5/2017   3:04:37 PM