Keypad Functions

Short press = <2 seconds 
Long press = >2 seconds

Calibration

1.  Rinse the probe in distilled water and gently dry with a 

clean tissue. Short press  /MEAS to turn on the tester.

2.  Long press CAL/

  to enter calibration mode;  

short press  /MEAS to exit.

3.  Dip the probe into 1413 µS/cm calibration 

solution. Stir gently, leave it to stand. Wait 

for the measurement stability icon (

) to 

appear and stay on the display (see Fig 1); 

then short press CAL/

  to complete the 1st 

calibration. Tester returns to measurement 

mode, and calibration icon “M” appears 

on bottom left side of display.

4.  Rinse probe in distilled water and dry it. Repeat steps  

#2 and #3 to complete 2nd calibration in 12.88 mS/cm 

calibration solution. Tester returns to measurement 

mode, and calibration icons “M” and “H” will appear 

on bottom left side of display.

Operating Manual

  

EcoTestr

 EC1 Pocket Tester

Setting the Parameters

When tester is off, long press  /MEAS to enter setup mode. 
Short press  /MEAS to switch from P1 to P2 to P3. Short 
press CAL/

  and parameter will flash, then short press  

/MEAS to choose desired parameter. Short press  

CAL/

  to confirm parameter selection. Long press  /MEAS 

to return to measurement mode.

Notes: 

•  Once the readings are greater than 2000 µS/cm, the unit 

will automatically become mS/cm.

•  Select “Yes” for P3 to restore the calibration to the 

 

theoretical values and parameter setting to original  

values. When tester’s calibration or measurement  

performs abnormally, this function can be adopted so 

the tester returns to factory default setting and then 

users can conduct calibration or take measurements 

again.  

Measurement

1.  Short press  /MEAS to turn on the tester. Rinse probe 

in distilled water and dry it gently with a clean tissue to 
remove excess water.

2.  Stir the probe in the sample solution gently, leave it to 

stand. Wait for the stability icon (

) to remain on, then 

take a reading.

3.  Rinse off the tester thoroughly in distilled water after 

each test. 

Notes: 

•  If this is the first-time use or the tester hasn’t been used 

for a long time, we recommend soaking the probe in 

12.88 mS/cm solution for 15 to 30 minutes to restore its 

sensitivity and speed up the response time.

•  The tester adopts 1413 µS/cm and 12.88 mS/cm standard 

 

calibration solutions. User can use 1 and 2 point calibra

-

tions as needed. For most circumstances, calibrating in 
1413 µS/cm to complete the 1st point calibration will 
meet testing requirements.

•  The tester has already been calibrated after manufac-

ture. User can use the tester immediately or can test it in 

the calibration solutions to test its accuracy. If error is 

large, calibrate the tester before using.

•  We recommend replacing the calibration solution after 

 

5 to 10 calibrations to maintain accuracy.

1433

1413

µS

Cond

Figure 1

Symbol

Menu setting

Selection

Factory  

default

P1

Select conductivity 

range*

Aut* – µS – 

mS

Aut

P2

Select temperature 

unit

°F – °C

°C

P3

Restore to  

factory default

No – Yes

No

*Aut = all ranges (0 to 199.9 µS/cm, 20 to 1999 µS/cm, 2.0 to 20.0 mS/cm)

1.  Short press to turn on; long press to  

turn off.

2.  When turned off, long press to enter  

setup mode.

3.  In mode setting, short press to change 

parameter.

1.  When turned on, long press to enter 

calibration mode.

2.  In calibration mode, short press to 

 

confirm calibration.

3.  In mode setting, short press to confirm 

parameter selection.

CAL

<

MEAS

1065O101_MAN_EC1_35634-06.indd   1

03/22/2018   3:49:52 PM