36
e700 User Manual
Figure 12.a
- BiLVL waveform w/o pressure support
TABLE 4
-
Default Ventilation Setting- BiLVL
PARAMETER
RANGE
DEFAULT
INFANT
CHILD
ADULT
Pi
(OFF, 4-50 cm H
2
O)
15
15
15
Rate
(5 - 60 BPM)
30
15
10
PEEP
(OFF, 4-20 cm H
2
O)
5
5
5
I:E ratio
(1:4 – 3:1)
1:2
1:2
1:2
Ti*
(0.2 - 9 sec)
0.66
1.33
2.0
PSV
(OFF, 4-35 cm H
2
O)
Off , unless
selected
Off , unless
selected
Off , unless
selected
Trig.
(1 -15 L/min)
3
3
3
Termination
20-80% of maximum
set fl ow
50%
50%
50%
O
2
%
(100% or 60% O
2
)
100
100
100
Mv max
(2 -40 L)
30
30
30
Mv min
(0.5 – 35 L)
0.5
0.5
0.5
P max.
10 -80 cm H
2
O
25
25
30
P min.
0-20 cm H
2
O
(during I time only)
3
3
3
Manual
Refer to Manual and
I-Hold section
ready
ready
ready
* Ti may be limited below its range depending on set I:E ratio and rate.