e600 User Manual

   

(15PL1010 - Rev.7  28_8_2017)    English version  

  

 

 

e600

 

Automatic Transport Ventilator 

 

User Manual 

 

 
 

Содержание e600

Страница 1: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version e600 Automatic Transport Ventilator User Manual...

Страница 2: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version...

Страница 3: ...5 1 Screen Layout 10 3 5 2 Live Monitoring Parameters 12 3 6 Symbols Notations 13 Chapter 4 Preparation For Use 14 4 1 Set up 14 4 1 1 Connecting electrical power supply 14 4 1 2 Installing replacing...

Страница 4: ...ning Disinfection 37 8 2 Charging the Battery 37 8 3 Ambient Air entrainment Filter 38 8 4 Servicing 38 Chapter 9 Technical Data 40 9 1 Specifications 40 9 2 Circuit Description 42 9 3 Battery and Pow...

Страница 5: ...onfined space will elevate oxygen levels The ventilation setting will be interrupted during battery replacement while the device is in operational mode and the external power supply is not connected O...

Страница 6: ...range specified in this manual may result in reduction and or failure in the ventilator s performance In extreme temperature conditions the effect is not noticeable in terms of delivered ventilations...

Страница 7: ...or must only be serviced by the manufacturer or its authorized service centers Device single use patient circuits and battery packs should be safely discarded in accordance with local state and instit...

Страница 8: ...emergency and transport ventilator designed for use in the pre hospital intra hospital inter hospital and transport settings It is intended for use with adult child infant patients with a tidal volume...

Страница 9: ...Resume button B Battery operation indicator I Lock button C External power indicator J Waveform selection button D Battery charging indicator K Cancel button E Manual Hold button L Confirmation indica...

Страница 10: ...After 4 seconds the ventilator will turn off and all ventilation will stop If the button is pressed and held for less than 4 seconds the ventilator will stay ON Warning At approximately 2 of full bat...

Страница 11: ...rms displayed on the screen 3 2 6 Screen brightness Pressing this button will switch between 4 different screens as follows 1 Light background with dark color text and waveform at 100 light intensity...

Страница 12: ...settings but no ventilation will occur unless Pause function is disabled 4 To cancel Pause function press Pause Resume button H again The Resume symbol will flash on the screen with the confirmation...

Страница 13: ......

Страница 14: ......

Страница 15: ...11 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version SIMV screen layout CPAP screen layout CPR screen layout...

Страница 16: ...c will be followed Rate BPM Rate BPM is the rate at which breaths are delivered in one minute It is the monitored breath rate calculated by measuring the time interval Tb in seconds between 2 breaths...

Страница 17: ...ote Offers useful tips to assist in the proper use of the equipment Confirmation Symbol Invalid Setting Symbol Conflict Setting Symbol Keep away from opened flames No smoking around ventilator IPX4 In...

Страница 18: ......

Страница 19: ...ylinder o Secure oxygen cylinders so they do not fall over o Keep away from excessive heat to avoid the risk of explosion o Do not grease or lubricate oxygen fittings cylinder valves and pressure redu...

Страница 20: ...turn on but with no ventilation at this point If the button is pressed and released for less than a second the ventilator will remain OFF Note Do not connect the patient circuit to a patient or the te...

Страница 21: ...linder pressure is above 650 PSI 45 bar otherwise replace with new oxygen cylinder Once pressurized turn off the oxygen cylinder and observe the pressure regulator gauge reading If pressure does not d...

Страница 22: ...ciated with yellow warning indicator BCI audible alarm associated with red warning indicator must activate in 15 seconds 6 Block patient output completely to activate Pmax alarm should be activated 7...

Страница 23: ...ilator is turned on 3 Solid Silhouettes start up figures Shown in Figure 6 will be displayed on the screen representing Infant Child and Adult patient sizes Each size comes with pre set parameters clo...

Страница 24: ...on 6 of the screen Figure 4 The user must confirm the selection by pressing the Control Selection Knob M Once confirmed the selected area will be highlighted with solid contrast background Figure 9 Na...

Страница 25: ...500 Mv L calculated values 3 0 3 7 5 0 PEEP cm H2O 5 5 5 P max cm H2O 25 25 30 P min cm H2O 3 3 3 Ti Sec 0 66 1 33 2 0 I E 1 2 1 2 1 2 O2 100 100 100 F trig L min 3 3 3 Figure 7 A C V mode screen w d...

Страница 26: ...e Note When switching between ventilation modes any shared parameter will be carried over and any new parameter will be set to the default 5 2 1 A C V Assist Control Ventilation In AC mode Figure 10 t...

Страница 27: ...E ratio 1 4 3 1 1 2 1 2 1 2 Ti 0 2 9 sec 0 66 1 33 2 0 PEEP OFF 4 20 cm H2O 5 5 5 Trig OFF 1 15 L min 3 3 3 O2 100 or 60 O2 100 100 100 P max 10 80 cm H2O 25 25 30 P min 0 20 cm H2O during I time onl...

Страница 28: ...start of inhalation phase beginning of Ti If no effort was detected during the trigger period the ventilator will initiate mandatory ventilation at the end of trigger window Figure 11 Should the patie...

Страница 29: ...25 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 11 SIMV waveform...

Страница 30: ...the ventilator does not trigger patient s spontaneous breathing for a period of time T APNEA set by the user The parameters of back up A C ventilation are defaulted to volume ventilation based on the...

Страница 31: ...27 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 12 CPAP ventilation waveform...

Страница 32: ...essions and ventilations is 30 2 Figure 13a The CPR mode for intubated patients consists of continuous compressions indicated by an audible prompt and visual animation at a rate of 100 compressions pe...

Страница 33: ...29 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version CPR for masked patients On screen chest compressing animation On screen ventilation animation Figure 13 a CPR waveform for masked patient...

Страница 34: ...30 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version CPR for intubated patients On screen intubated CPR animation Figure 13 b CPR waveform for intubated patient...

Страница 35: ...31 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version 5 3 Turning the Ventilator Off Press and hold the ON OFF button for 4 seconds the Ventilator will turn OFF...

Страница 36: ...pply to the ventilator b Disconnect gas supply hose c Disconnect patient circuit from the output connector d Unplug the power cable from mains if no charging is required e Clean and disinfect accordin...

Страница 37: ...tinue to flash until the cause of the problem is resolved During alarm silence should a new alarm develop the Alarm Silence function will continue and only the new alarm will be shown flashing on the...

Страница 38: ...ng symbol Red Two bursts with five pulses Low Oxygen 40 PSI Medium Flashing symbol Yellow One burst with three pulses High input pressure 90 PSI High Flashing symbol Red Two bursts with five pulses AP...

Страница 39: ...le 7 2 Battery charging status 1 Full Capacity No Alarm 2 95 of full capacity No Alarm 3 90 of full capacity No Alarm 4 80 of full capacity No Alarm 5 65 of full capacity No Alarm Warning At approxima...

Страница 40: ...depends on the severity of the failure Green color LED Continuous when unit is connected to AC power source during both ON and OFF phases Orange color LED Continuous when unit is charging and off when...

Страница 41: ...ent circuit Using a wet or damp filter may result in inaccurate parameters and potentially damage the ventilator Warning Risk of Explosion Cleaning agents containing alcohol or grease become flammable...

Страница 42: ...Always keep the ambient air entrainment port clear of obstructions Always replace the filter after use Warning Avoid particulate and or gaseous pollutants in the ambient air The entrainment of polluta...

Страница 43: ...h use applications Any malfunction unit should be returned to the manufacturer or an authorized service center since this product is not designed for field disassembly or service Unauthorized repairs...

Страница 44: ...nspiratory hold Maximum Inspiratory hold time 6 sec I E Ratio 1 4 3 1 20 PEEP cm H2O 0 4 20 10 or 2 cm H2O CPAP cm H2O 4 20 10 or 2 cm H2O O2 60 or 100 15 Pmax cm H2O 10 80 10 or 2 cm H2O Pmin cm H2O...

Страница 45: ...th mask Dead space of patient valve with elbow Approx 35 ml Compliance disposable hose system 16 6 ml kPa Resistance of Patient Hose system Inhalation and Exhalation Less than 6 cmH2O at 60 l min Less...

Страница 46: ......

Страница 47: ...behaviour of the ventilator will not be affected while the battery is charging AC DC Power Supply Model PMP90 13 2 01CV0105 Input Voltage 100 240 VAC Input Current 1 06 0 45 A Input Frequency 47 63 Hz...

Страница 48: ......

Страница 49: ...ower frequency 50 60 Hz magnetic field IEC61000 4 8 3 A m 30 A m Conducted RF IEC61000 4 6 3Vrms 50kHz to 80 MHz outside ISM bands 3 Vrms 10Vrms 150kHz to 80 MHz in ISM bands 10 Vrms Radiated RF IEC61...

Страница 50: ......

Страница 51: ...patient ventilation circuit check tube position adjust Pmax alarm value Apnea spontaneous breathing failed or disconnection faulty sensor Switch to Assist Control ventilation ensure connections tight...

Страница 52: ...er minute I E Ratio of inspiration time to expiration time LED Light Emitting Diode Mve Exhaled Minute Volume Mv Minute volume O2 Percentage of Oxygen inspired Paw AV Average airway pressure Paw peak...

Страница 53: ...wer supply Cord Canada and the US 01CV0106 5 Battery Pack 01CV9100 6 e Ventilator External Power Supply 01CV0105 7 SMART MOUNT Mounting Bracket for Road ambulance 01EV7035 8 FIXED MOUNT Mounting Brack...

Страница 54: ...s deemed to be out of specification must be returned to O Two Medical Technologies Inc or one of its approved Service Centers for service and or repair We recommend that a service contract be obtained...

Страница 55: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version...

Страница 56: ...SN ____________ O Two Medical Technologies Inc 45A Armthorpe Road Brampton Ontario Canada L6T 5M4 Office 1 905 792 OTWO 6896 N A Toll Free 1 800 387 3405 Fax 1 905 799 1339 Email resuscitation otwo co...