43
e600 User Manual
(15PL1010 - Rev.7 28_8_2017) English version
9.3 Battery & Power Supply
Battery Pack
Battery Cell Type
Rechargeable Lithium Ion Cell
Type 4ICR19/65-3
Nominal Capacity
7500 mAh, Min 111 Wh
Nominal Voltage
14.8 V
Max. Charging
Current
3750 mA
Max. Charging
Voltage
16.8 V
±
0.1 V
Dimension
144 x 62 x 42
Weight 642
g
Test specification
Meet requirements of
IEC62133:2002
Note
The behaviour of the ventilator will not be affected while the battery is
charging.
AC/DC Power Supply
Model
PMP90-13-2 01CV0105
Input Voltage
100 - 240 VAC
Input Current
1.06 - 0.45 A
Input Frequency
47 - 63 Hz
Output Voltage
19 VDC
Output Current
4.74 A Maximum
DC Output Plug
2.5 x 5.5 x 11 mm
DC Output Cable
Length
6’
AC Power Cord
Length
6’
Weight 642
g
EMC performance
Meet requirements of IEC60601-1-2
Safety Standards
Meet requirements of IEC60601-
1:2005
Note
Upon disconnecting AC Power supply, Ventilator will automatically switch to
Battery operation without affecting ventilator behaviour.
Содержание e Series
Страница 2: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...
Страница 13: ......
Страница 14: ......
Страница 18: ......
Страница 29: ...25 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 11 SIMV waveform ...
Страница 31: ...27 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 12 CPAP ventilation waveform ...
Страница 46: ......
Страница 48: ......
Страница 50: ......
Страница 55: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...