Along with the contents of the shipping carton you will require the 
following    items to enable you to undertake the pre-use functional 
check: 

 

[1] 

Full breathing air or medical grade oxygen cylinder. 

 

[2] 

Air or oxygen regulator with a 50 PSI outlet and the correct DISS or 
quick connector  for  the  gas  being  used.  The  regulator  must  be 
able  to  deliver  an  output  flowrate  of  >120  L/min  at  no  less  than 
45 PSI (3.1 Bar). 

 

[3] 

Calibrated  Test  Lung. 

 

Having  connected  the  supply  hose  to  the  regulator,  turn  on  the  gas 
supply.  Using  a  mild  soap  solution,  spray  the  input  connection  to  the 
resuscitator  to  check  for  leaks.  If  any  leak  is  present,  tighten  the 
connection “by hand” (Do not use a wrench) and re-test. 

 

Once  no  leaks  are  found,  connect  the  test  lung  to  the  15/22mm  patient 
connector on the  resuscitator.  If  the  leak  cannot be  stopped,  check  and 
replace the sealing “O” ring on the hose connection to the resuscitator. If 
the leak persists, replace the hose. 

 
 

 

The  following  features  can  be  individually  tested  during  the  Pre-use 

Functional Check: 

 

 

[1]        Peak Airway Pressure and Audible Alarm.  

[2]        Ventilation Frequency and Tidal Volume.  

[3]        Manual Ventilation. 

[4]        Demand Valve Function and Automatic Circuit Shut Off.  

[5]        Bleed Flow. 

 

 

 

The supply hose provided is attached to the oxygen inlet on the rear of 
the resuscitator and is tightened “finger tight” (fig 1). 
 

         The use of excessive force in tightening the supply hose 

may damage the seal and /or thread. 

 

The  facemask  is  attached  to  the  patient  connection  port  by  simply 
pushing the mask onto the 22mm taper. 

 

 

Manual Button 

 

 
 
 
 
 
 

Attach supply 

hose  here 

 
 

 

 

 

 

The 

®

has  a  Manually  Actuated, 

Automatic Ventilation Override Button (Manual Button)  with a 20 second delay 

to automatic cycling re-start to assist in the timing of ventilations in conjunction 

with external cardiac compressions. 

 

By using the Manual Button”, the operation of the ventilator can be easily 

timed with the chest compressions  so as  to avoid the potential problem of 

the  aspiration  of  stomach  contents  due  to  gastric  distension  which  may 

occur if overlap of chest compression  and  inflation  occurs.    (It  has  been 

shown in some studies that, in  patients that are intubated, this overlap of 

compression and inflation may increase cardiac output without the danger 

of gastric distension.) The flowrate provided is  equivalent  to the preset 

automatic flowrate. 

 

Содержание CAREvent CA-G05

Страница 1: ...dible Alarm 2 Ventilation Frequency and Tidal Volume 3 Manual Ventilation 4 Demand Valve Function and Automatic Circuit Shut Off 5 Bleed Flow The supply hose provided is attached to the oxygen inlet on the rear of the resuscitator and is tightened finger tight fig 1 The use of excessive force in tightening the supply hose may damage the seal and or thread The facemask is attached to the patient co...

Страница 2: ...serve the patient s chest movements If there is any leak from around the mask or any obstruction in the patients airway blow off valve will operate reposition patients head and adjust mask and hand position to ensure a good airway and mask to face seal If the patient is intubated or if the patient is to be intubated following mask ventilation remove the face mask from the 22mm connector and attach...

Страница 3: ...ze cylinder containing 415 Litres of air or oxygen Should the patient vomit into the mask during resuscitation the following steps should be followed to clear the foreign material 1 Remove the mask from the patient s face and clear any foreign material from the patients airway Depress the manual button or allow the resuscitator to cycle automatically for a few breaths to clear the mask and valve o...

Страница 4: ...accessories for correct function should be undertaken by a responsible member of staff on a routine basis This check is to ensure that all of the accessories and resuscitator components are present the air or oxygen cylinder is full and that the resuscitator is in working order Regulator working pressure suction if equipped and ventilator limiting pressures should be checked at least every six mon...

Страница 5: ...tilator support in pulmonary resuscitation in emergency medical and rescue environment where toxic gas may be present The ventilator is suitable for use in Pulmonary resuscitation during respiratory and or cardiac arrest Short term ventilatory support in the confined space rescue in IDLH Immedi ately Dangerous to Life and Health environment pre hospital Intra hospital inter hospital and air ambula...

Страница 6: ...ssemble unit connect to an air or oxygen supply to check operation prior to packaging for emergency use 17MP9039 Reusable patient valve swivel housing 17MP1528 Reusable Patient Valve Diaphragm 01FG6500 Mask Retaining Insert 01FG6501 Mask Insert Extraction Tool 01FG6502 Head Harness System 17MP7010 Single Use PEEP Valve 01FV4314 Conversion Hose O2 Air 1 1 1 Introduction 1 1 2 Warranty Information 1...

Страница 7: ... ...

Отзывы: