FLOW CONTROLLERS

FOR USE WITH MEDICAL GASES

INTENDED USE:

The O-Two Medical Technologies “Therapy Flow Controller” is intended for the delivery of adjustable 

therapy fl ow rates (medical oxygen or medical air) to a patient.

DESCRIPTION:

The O-Two “Therapy Flow Controller” is a therapy fl ow administration device which is capable of 

delivering a controllable continuous fl ow of medical gas (oxygen or air) within the specifi ed range.

Connected to medical gas supply system, O-Two “Therapy Flow Controller” can deliver the desired 

therapy fl ow by adjusting fl ow-selecting control.  

Rev. 7 Feb 2020

O-TWO MEDICAL TECHNOLOGIES INC.

45A Armthorpe Road, Brampton, Ontario, Canada L6T 5M4

Tel. (905) 792-6896   Fax (905) 799-1339 

www.otwo.com

REF

01FC

 SAFETY WARNINGS       

1. 

It is imperative that the O-Two Therapy Flow Controller only be operated in accordance with the 

instructions as described in this manual by clinically trained and authorized personnel. Do not use the 

Flow Controller for driving any medical equipment.

2. 

Using oil, grease or other petroleum-based substances with oxygen and oxygen mixtures may cause 

fi re or explosion. Never handle Flow Controllers with oily or greasy hands or gloves. Do not use or store 

oxygen equipment near excessive heat or open fl ame.  

3. 

Make sure that the inlet fi ttings on fl ow controllers are properly mated for the gas intended.  Never 

attempt to force an incompatible connection or change the connectors. Never interchange fl ow 

controllers, hoses, or other equipment with similar equipment intended for use with other gases.

4.

 Disconnect the fl ow controller from a pressure source and release all gas from the fl ow controller 

after each use. 

5.

 Do not stand in front of a fl ow controller outlet when opening the pressure source in case foreign 

particles are present which could cause a hazardous malfunction of the fl ow controller. Never occlude 

the outlet to test for the presence of pressure.

6.

 The fl ow controller doesn’t indicate that a fl ow is occurring. 

7.

 Setting the fl ow-selecting knob between adjacent settings may cause the output fl ow to cease! 

8.

 Varying inlet pressure outside the specifi ed operating inlet pressure has signifi cant impact on the 

accuracy of output therapy fl ow rate!

9.

 Presence of outlet resistance may result in decrease of the fl ow rates delivered. 

10.

 Ambient temperature variations within the specifi ed operating temperature range have no eff ect 

on the accuracy of fl ow rate.

11. 

Oxygen therapy may be critical treatment. The application of oxygen therapy using this fl ow 

controller should be made in strict accordance with the prescription and instructions of a physician.

12. 

Never disassemble a fl ow controller under pressure! Disassembly and assembly of fl ow controllers shall 

be performed only by O-Two authorized and trained personnel. Unauthorized disassembly will void the 

warranty. Lubricant, if used in the maintenance of the fl ow controller, shall be compatible with the intended 

medical gases. Use of lubricants not recommended by the manufacturer may cause fi re or explosion.

SPECIFICATIONS:

GAS

Medical Oxygen, Medical Air

OPERATING INLET PRESSURE

50 to 55 PSI

THERAPY FLOWRATE RANGE

0 – 6 l/min, 0 – 8 l/min, 0 – 15 l/min, 0 – 25 l/min

STORAGE TEMPERATURE

From -28°C to +60°C

OPERATING TEMPERATURE

From -18°C to +50°C

WARRANTY:

The O-Two Therapy Flow Controllers for Use with Medical Gases are manufactured from the 
fi nest quality materials.  Each individual part is subject to strict quality control tests to ensure 
exceptionally high standards.  The manufacturer warrants to the purchaser of the O-Two 
Flow Controller that its component parts are free from defects in material and workmanship 
for a period of fi ve (5) years from the date of purchase. The manufacturer will replace and/
or repair all parts of the Flow Controller at its option for fi ve years from the date of purchase 
at no cost to the purchaser, upon the notifi cation of the defects, in writing by the purchaser 
and confi rmation of said defect by the manufacturer.  All shipping costs shall be borne by 
the purchaser. The manufacturer shall be liable under this warranty only if the Flow Controller 
and its parts have been used and serviced in the normal manner described in the instruction 
manual.  This warranty does not cover failure resulting from alterations, accidental damage, 
misuse or operation not in accordance with the instructions provided. Wear items are not 
covered hereunder. This warranty is null and void if unauthorized repairs are attempted or 
performed. There are no other expressed or implied warranties.  This warranty gives no 
specifi c legal rights.