Invu by Nuvo™
User Manual
Maintenance
NV-UMLS-0012
65
Item
Inspection
2- Detachable (bio-potential) sensors
Check all
8
detachable sensors for any damage or
tear
3- Acoustic sensors
Check all
4
acoustic sensors for any damage or
tear
Figure 7-3. INVU Accessories
Item
Inspection
1- ECG gel (Spectra 360®)
Make sure the ECG
gel packing isn’t
damaged or
are not intact
2- Cleaning Isopropyl Alcohol Spray
(Safetec)
Make sure the cleaning spray isn’t
damaged or
are not intact
3- Quick User Guide
Make sure Quick User Guide provided in the
personal kit is legible for reading.
4- Medical grade power supply and
charging cable
Check provided power supply for any damage. Do
not use any other power supply but the one
provided with the kit.
7.2
Cleaning the
INVU Sensor Band™
Cleaning Before Use
Wipe the
INVU Sensor Band™
with CaviWipes™ wipes or equivalent, removing all gel
residues. If an excessive amount of gel was applied on the sensor, wipe the excess gel
that overflows the cavity with CaviWipes™ wipes or equivalent.
Clean the
INVU Sensor Band™
with a hospital-grade alcohol disinfectant spray.
Let the band parts air-dry for 10 min.
Wipe the band and sensors with a dry lint-free cloth.
Cleaning After Use
It is very important to clean the band immediately after completing a monitoring session.
Avoid accumulation and drying of gel on the band and its sensors.