Do not use corrosive materials which may be harmful for the device integrity, rotor and
accessories.
Do not use rotors and accessories with corrosion and mechanical damages.
Mains supply should be appropriate to power of the device and grounded.
Use tubes whose sizes are suitable to the rotor and accessories.
Tubes which are used in the centrifuge should not be deformed by the effect of the centrifuge
force.
Use glass tubes to balance, if glass tubes are used. Use plastics tubes to balance, if plastics
tubes are used.
Do not start the device unless tubes are in balance.
Imbalance loading may cause mixing the samples, broken tubes, and damages on the rotor
and motor shaft.
SECTION 3
SYMBOLS
Symbol in the operating instructions:
Attention, general hazard area.
This symbol refers to safety relevant warnings and indicates possibly dangerous
situations.
The non-adherence to these warnings can lead to material damage and injury to
personal.
Symbol in the operating instructions:
This symbol refers to important circumstances.
SECTION 4
INSTALLATION PROSEDURE
4.1 LIFTING AND TRANSPORT
All lifting and transport must be carried out by using proper handling equipment. The instrument
must be supported from underneath and never turned over.
4.2 UNPACKING
Remove the packing cardboard box and the second nylon packing around the centrifuge. The
below written are provided with the instrument, please check them;
1 ea. U
ser’s manual
1 ea. Warranty certification
1 ea. Power cable
1 ea. Manual lid opening tool