NUVE FR 290 Скачать руководство пользователя страница 6

 

1.

 

INTRODUCTION 

1.1.

 

USE AND FUNCTION 

FR  290,  FR  490  and  FR  590  Freezers  are  designed  for  the  storage  of  the  samples  at  low 
temperature  and  it  is  used  in  the  hospitals,  blood  banks,  epidemic  diseases  research  center, 
research laboratories in the universities, genetic engineering, bio-pharmaceutical industry. 

FR series Freezers are capable of cooling down to set points between -20°C and -45°C and keep 
temperature stable within given tolerances. Temperature in the chamber is displayed digitally. It 
is possible to set the working temperature in the range of -20°C to -45°C with 1°C sensitivity. FR 
series freezers are safely operated via 

N-SmArt™

 microprocessor control system. 

The refrigerant gas passes through evaporators placed under each shelf, bottom of the chamber 
and top of the chamber; thus, each compartment of freezer is cooled directly. FR series provides 
a  stronger  and  more  rapid  cooling  with  its 

DirectFREEZE™

 

technology.  The  refrigeration  is 

provided by non-explosive refrigerant gas which does not include CFC. 

The power switch is equipped with a key lock to prevent unauthorized usage. Door handle on FR 
series can be opened easily. Moreover, silicon gasket placed on the door protects air leakage.  

FR  series  freezers  has  audible  and  visual  alarm  systems  for  high  and  low  temperature 
conditions,  open  door,  power  failure,  temperature  sensor  failure  and  other  faults  associated 
with  the  microprocessor  control  system.  The  alarm  system  is  fed  by  an  automatically  re-
chargeable  battery.  FR  series  store  last  50  failures  in  the  memory  and  these  errors  can  be 
followed  on  the  display.  Standard  Ethernet  connection  feature  enables  email  sending  to  five 
different  e-mail  addresses  in  case  of  any  error.  FR  series  offer  GSM  module  as  an  option  and 
SMS can be sent to five different telephone numbers in case of error. 

In case of electricity interruption, the chamber temperature is measured and displayed, as well. 
Remote  alarm  outlet  and  the  outlet  for  connection  to  central  alarm  system  are  offered  as 
standard.  

FR series have internal memory which can store temperature values for 10 years. In addition to 
internal  memory,  temperature  values  can  be  recorded  in  USB  stick.    FR  series  have  a  weekly 
circular chart recorder as an option. The temperature in the chamber is recorded by this chart 
recorder operated by the backup battery. Chart recorder is equipped with a key lock to prevent 
unauthorized usage. 

FR 290, FR 490, FR 590 Freezers are manufactured according to the following standards: 
EN  60601-1,  EN  61010-1,  EN  61000-6-3,  EN  62366,  EN  980,  EN  60601-1-4,  EN  60601-1-2,           
EN 60601-1-6 and EN ISO 14971, EN 50419. 

This device is in compliance with WEEE Regulation. 

If  the  warnings  mentioned  in  this  manual  are  not  considered,  NUVE  will  not  be  responsible 
from their results. 

Содержание FR 290

Страница 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A FR 290 FR 490 FR 590 DEEP FREEZER USER S MANUAL 1984 Z14 K 25 275 Rev No 08 Rev Date 01 2016...

Страница 2: ...ate damaged instrument 4 do not operate the instrument with a damaged cable 5 do not move the instrument during operation In case of a problem contact your N ve agent for an authorized service or main...

Страница 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...

Страница 4: ...5 5 POSITIONING 10 5 6 GENERAL PRESENTATION 11 5 6 1 CHAMBER DIMENSIONS 12 5 7 CONTROL PANEL 12 5 8 PRIOR TO OPERATION 14 6 OPERATING PRINCIPLES 14 6 1 PROGRAMMING 14 6 1 1 OPERATING MODE 14 6 1 2 SET...

Страница 5: ...4 10 1 AlerText GSM MODULE 24 10 2 N veWARN REMOTE ALARM SYSTEM 25 10 2 1 MOUNTING REMOTE ALARM 25 10 3 CHART RECORDER 26 11 ELECTRICAL CIRCUIT DIAGRAM 27 11 1 FR 290 FR 490 AND FR 590 ELECTRICAL CIRC...

Страница 6: ...emperature conditions open door power failure temperature sensor failure and other faults associated with the microprocessor control system The alarm system is fed by an automatically re chargeable ba...

Страница 7: ...mpartments internal isolation doors 3 3 4 4 Control System N SmArt Programmable microprocessor system Temperature Indicator Colorful LCD Temperature Recorder Circular Battery operated 7 day chart with...

Страница 8: ...mm A 08 159 9 x 9 Divider for up to 13 6 mm tubes Divider height 65 mm A 08 196 Rack with two drawers FR 590 A 08 197 Rack with three drawers FR 590 A 08 198 Rack for 133 x 133 x 75 mm cryoboxes FR 59...

Страница 9: ...instructions This symbol refers to important circumstances Labels on the device Earthed Wall Sockets Remote alarm output Normally closed contact output FR 290 490 590 Fuses 2 x 13 A 5 INSTALLATION 5 1...

Страница 10: ...hat the supplied mains matches the required power ratings which are written on the name plate of the instrument located at the back of the instrument Always plug in the instrument to correctly grounde...

Страница 11: ...11 5 6 GENERAL PRESENTATION Figure 1 Figure 2 CONDENSER FILTER USB PORT 1 USB PORT 2 ETHERNET RS 232 CO2 BACK UP OUTPUT FUSES POWER CORD POWER LED NC CONTACT OUTPUT REMOTE ALARM OUTPUT POWER KEY...

Страница 12: ...mpartments different from FR 490 and FR 590 Figure 3 5 7 CONTROL PANEL Figure 4 The functions of F1 F2 F3 and F4 buttons depend on the meaning of the corresponding symbol appearing on the display The...

Страница 13: ...return the previous page or exit from the page by pushing the button corresponding to this symbol This symbol denotes the value increase key You can increase the value by pushing the button correspon...

Страница 14: ...tely in 3 hours if the chamber is not loaded This period depends on the ambient temperature 6 OPERATING PRINCIPLES 6 1 PROGRAMMING At most 70 of the surface area of the shelves should be loaded in ord...

Страница 15: ...2 SETTINGS Settings page appears when settings submenu is selected from the menu page by the value increase F1 and decrease F2 keys The required submenu is chosen by using the value increase and decr...

Страница 16: ...shing the value increase F1 and decrease F2 If the option of open is selected activation time of screen saver is adjusted on screen saver time in minutes If you want to exit this page without any chan...

Страница 17: ...ng the value increase and decrease keys Push the enter key F4 to access the table Temperature values corresponding to the selected date are shown on three pages Push the value increase F1 and decrease...

Страница 18: ...age there are three options Copy History Instant Record and Copy Failure History Copy History As USB stick is connected to the device all data in the internal memory is transferred to USB stick when C...

Страница 19: ...crease F1 and decrease F2 keys The page on the left appears when enter key F4 is pushed The phone numbers can be entered in this page and SMS notifications are delivered to these phone numbers in case...

Страница 20: ...the e mail addresses E Mails is selected on the e mail page by the value increase F1 and decrease F2 keys The page on the left appears when enter key F4 is pushed The e mail addresses can be typed in...

Страница 21: ...is used for changes from one character of address to another 6 1 6 SET TEMPERATURE When the setting key F2 is pushed in the operation mode password page appears This password is the user s password Th...

Страница 22: ...enser filter should be cleaned once in a month To access condenser filter pull out the front panel of the compressor compartment See Figure 5 Take out the filter and clean the filter with compressed a...

Страница 23: ...st be disposed off according to the regulations for the disposal of electronic components Batteries power packs and similar power source have to be dismounted from electric electronic parts and dispos...

Страница 24: ...se contact with an authorized N ve agent to seek technical help 9 2 FUSE REPLACEMENT The fuses shall always be replaced by authorized personnel 10 OPTIONS 10 1 AlerText GSM MODULE In case of error FR...

Страница 25: ...ppears in the error history If the GSM module is connected and cannot send messages SMS error code appears in the error history Modem and SMS errors do not appear when SMS submenu on the SMS page is s...

Страница 26: ...unit and take it out Put on the battery supplied with device to the hole at the back of the unit by paying attention to and poles Always use leak protected high quality batteries Put the battery unit...

Страница 27: ...27 11 ELECTRICAL CIRCUIT DIAGRAM 11 1 FR 290 FR 490 AND FR 590 ELECTRICAL CIRCUIT DIAGRAM...

Отзывы: