5.8.

 

PRIOR TO OPERATION 

 

Plug-in the instrument to correctly grounded sockets. 

 

Insert the key to the power switch lock at the rear panel. 

 

Power the freezer on by rotating the key clockwise. 

 

Observe that the microprocessor control system is activated. 

 

Learn the function of the control panel (See Section 5.7.) 

 

The  freezer  starts  automatically  according  to  the  last  set  temperature.  You  should  set  the 
temperature value according to your need (See Section 6.1.). 

 

When freezer starts to operate, the temperature of the chamber will appear on the control 
panel and audible and visual alarm will be activated. Push alarm mute key (F2) to mute the 
audible  alarm.  Temperature  of  chamber  appears  on  the  control  panel  during  cooling 
process.  

 

 

FR  series  freezers  will  reach  -41°C

 

approximately  in

 

3  hours  if  the  chamber  is  not 

loaded. (This period depends on the ambient temperature.) 

6.

 

OPERATING PRINCIPLES 

6.1.

 

PROGRAMMING 

 

At most 70% of the surface area of the shelves should be loaded in order to obtain a 
uniform temperature distribution. 

 

 

The  temperature  alarm  range  can  be  adjusted  between  5

C  and  10

C.  Red  colored 

screen meaning warning will be displayed when freezer is operated for the first time 
and audible alarm will be activated. Push the alarm mute key (F2) to mute the audible 
alarm. The red screen turns into the blue when temperature in the chamber is in the 
temperature alarm range.  

6.1.1.

 

OPERATING MODE 

 

When  the  freezer  is  powered  on,  it  starts  to 
operate  automatically  according  to  the  last  set 
values. The screen on the left is operating mode 
screen. 

Set 

temperature 

and 

chamber 

temperature values appear in this screen. Current 
time and date are shown at the top line. “

” will 

appear and flash on this screen when the freezer 
starts cooling process. 

 

Содержание FR 290

Страница 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A FR 290 FR 490 FR 590 DEEP FREEZER USER S MANUAL 1984 Z14 K 25 275 Rev No 08 Rev Date 01 2016...

Страница 2: ...ate damaged instrument 4 do not operate the instrument with a damaged cable 5 do not move the instrument during operation In case of a problem contact your N ve agent for an authorized service or main...

Страница 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...

Страница 4: ...5 5 POSITIONING 10 5 6 GENERAL PRESENTATION 11 5 6 1 CHAMBER DIMENSIONS 12 5 7 CONTROL PANEL 12 5 8 PRIOR TO OPERATION 14 6 OPERATING PRINCIPLES 14 6 1 PROGRAMMING 14 6 1 1 OPERATING MODE 14 6 1 2 SET...

Страница 5: ...4 10 1 AlerText GSM MODULE 24 10 2 N veWARN REMOTE ALARM SYSTEM 25 10 2 1 MOUNTING REMOTE ALARM 25 10 3 CHART RECORDER 26 11 ELECTRICAL CIRCUIT DIAGRAM 27 11 1 FR 290 FR 490 AND FR 590 ELECTRICAL CIRC...

Страница 6: ...emperature conditions open door power failure temperature sensor failure and other faults associated with the microprocessor control system The alarm system is fed by an automatically re chargeable ba...

Страница 7: ...mpartments internal isolation doors 3 3 4 4 Control System N SmArt Programmable microprocessor system Temperature Indicator Colorful LCD Temperature Recorder Circular Battery operated 7 day chart with...

Страница 8: ...mm A 08 159 9 x 9 Divider for up to 13 6 mm tubes Divider height 65 mm A 08 196 Rack with two drawers FR 590 A 08 197 Rack with three drawers FR 590 A 08 198 Rack for 133 x 133 x 75 mm cryoboxes FR 59...

Страница 9: ...instructions This symbol refers to important circumstances Labels on the device Earthed Wall Sockets Remote alarm output Normally closed contact output FR 290 490 590 Fuses 2 x 13 A 5 INSTALLATION 5 1...

Страница 10: ...hat the supplied mains matches the required power ratings which are written on the name plate of the instrument located at the back of the instrument Always plug in the instrument to correctly grounde...

Страница 11: ...11 5 6 GENERAL PRESENTATION Figure 1 Figure 2 CONDENSER FILTER USB PORT 1 USB PORT 2 ETHERNET RS 232 CO2 BACK UP OUTPUT FUSES POWER CORD POWER LED NC CONTACT OUTPUT REMOTE ALARM OUTPUT POWER KEY...

Страница 12: ...mpartments different from FR 490 and FR 590 Figure 3 5 7 CONTROL PANEL Figure 4 The functions of F1 F2 F3 and F4 buttons depend on the meaning of the corresponding symbol appearing on the display The...

Страница 13: ...return the previous page or exit from the page by pushing the button corresponding to this symbol This symbol denotes the value increase key You can increase the value by pushing the button correspon...

Страница 14: ...tely in 3 hours if the chamber is not loaded This period depends on the ambient temperature 6 OPERATING PRINCIPLES 6 1 PROGRAMMING At most 70 of the surface area of the shelves should be loaded in ord...

Страница 15: ...2 SETTINGS Settings page appears when settings submenu is selected from the menu page by the value increase F1 and decrease F2 keys The required submenu is chosen by using the value increase and decr...

Страница 16: ...shing the value increase F1 and decrease F2 If the option of open is selected activation time of screen saver is adjusted on screen saver time in minutes If you want to exit this page without any chan...

Страница 17: ...ng the value increase and decrease keys Push the enter key F4 to access the table Temperature values corresponding to the selected date are shown on three pages Push the value increase F1 and decrease...

Страница 18: ...age there are three options Copy History Instant Record and Copy Failure History Copy History As USB stick is connected to the device all data in the internal memory is transferred to USB stick when C...

Страница 19: ...crease F1 and decrease F2 keys The page on the left appears when enter key F4 is pushed The phone numbers can be entered in this page and SMS notifications are delivered to these phone numbers in case...

Страница 20: ...the e mail addresses E Mails is selected on the e mail page by the value increase F1 and decrease F2 keys The page on the left appears when enter key F4 is pushed The e mail addresses can be typed in...

Страница 21: ...is used for changes from one character of address to another 6 1 6 SET TEMPERATURE When the setting key F2 is pushed in the operation mode password page appears This password is the user s password Th...

Страница 22: ...enser filter should be cleaned once in a month To access condenser filter pull out the front panel of the compressor compartment See Figure 5 Take out the filter and clean the filter with compressed a...

Страница 23: ...st be disposed off according to the regulations for the disposal of electronic components Batteries power packs and similar power source have to be dismounted from electric electronic parts and dispos...

Страница 24: ...se contact with an authorized N ve agent to seek technical help 9 2 FUSE REPLACEMENT The fuses shall always be replaced by authorized personnel 10 OPTIONS 10 1 AlerText GSM MODULE In case of error FR...

Страница 25: ...ppears in the error history If the GSM module is connected and cannot send messages SMS error code appears in the error history Modem and SMS errors do not appear when SMS submenu on the SMS page is s...

Страница 26: ...unit and take it out Put on the battery supplied with device to the hole at the back of the unit by paying attention to and poles Always use leak protected high quality batteries Put the battery unit...

Страница 27: ...27 11 ELECTRICAL CIRCUIT DIAGRAM 11 1 FR 290 FR 490 AND FR 590 ELECTRICAL CIRCUIT DIAGRAM...

Отзывы: