The Pro CO can be used to analyze a regulated gas sample flow, the contents of a gas cylinder,
or the flow from a regulator:
•
If analyzing a gas flow, the Sample Flow Method is the preferred method. The flow rate must
equal 1.0-5.0 L/min at atmospheric pressure (1 bar). To produce this flow, a Flow Restrictor
and Regulator may be required. Contact Nuvair if you need assistance.
11.7 Pro CO Sample Flow Method of Checking Calibration
Nuvair offers CO gas testing kits see addendum for more information.
Testing Gas must not contain Helium for calibration to succeed.
Step 3. Allow 15
Seconds for Display
Reading to Stabilize
Step 4. Record
Reading while Gas is
Flowing
Flow Restrictor/
Regulator
Assembly
10 PPM CO
Calibration Gas
Canister
Nuvair
Calibration Gas
Specs:
CO 10 ppm
CO
2
1000 ppm
Nitrogen Balance