background image

OWNERS MANUAL

OWNERS MANUAL

Содержание GrowBoss

Страница 1: ...OWNERS MANUAL OWNERS MANUAL ...

Страница 2: ...ng pH Probes 8 Every Cleaning 8 Every Third Cleaning 8 Storing the pH Probe 9 If the probe has been allowed to dry out 9 Caring for Calibration Solutions 9 Storage 9 Replacement 9 Specifications 10 Abilities 10 Operational Environment 10 Troubleshooting Guide 11 Problem No Display 11 Problem pH Probe Inaccurate 11 Problem Temperature Inaccurate 11 Problem Nutrient reading is inaccurate 11 Problem ...

Страница 3: ...x Total Recall Min Max Indicators On Off Power up External Connections Nutrient Scale pH BNC USB Nutrient Power Supply Display Panel Plant Friendly High vis LED pH 7 Calibration trim pot pH 4 10 Min Max Nutrient Calibration trim pot Nutrient Scale indicator 5 67 EC On Off Power up ...

Страница 4: ...ur dealer immediately if there is damage to any of the above contents or if a component is missing 4 Before You Begin Your Nutradip GrowBoss is engineered to provide years of dependable service however as with any monitoring equipment it does require periodic calibration and maintenance In order to optimize meter longevity and performance we strongly recommend that you thoroughly read this manual ...

Страница 5: ...plug firmly into the GrowBoss Preparing the pH Probe 1 Remove the protective liquid filled cap from the pH probe Do not touch the pH probe bulb Put cap in a safe place The pH buffer solution is required for storage of the pH probe 2 Gently shake the probe downward to remove any air bubbles which may be trapped within the glass bulb Your Nutradip Growboss has been personally calibrated by a trained...

Страница 6: ...0 8 Place the probe in the clear tap water cup and agitate for 10 seconds Let stand for 60 seconds 6 Calibration Calibration should be done on your Nutradip GrowBoss at least once a month whenever a probe is cleaned when a new cycle is about to begin or when the GrowBoss has been taken out of storage Over time all probes naturally degrade due to environmental stresses and calibration ensures that ...

Страница 7: ...ed setting and follow the above procedure Refer to the conversion graph on page 10 7 9 Place the probe in the pH 4 0 solution and agitate for 10 seconds Let stand for 60 seconds 10 Rotate the 4 0 calibration screw counter clockwise very slowly and gently until the display increases to 4 0 11 Place the probe in the clear tap water cup and agitate for 10 seconds Let stand for 60 seconds 12 Repeat st...

Страница 8: ...insed with tap water to remove all traces of detergent and left overnight in pH 4 storage solution 8 Care Cleaning Regular care and cleaning of the Nutradip GrowBoss and its probes is the best way of ensuring that your device has a long and trouble free life Caring for pH Probes The pH Probe is the only component of the GrowBoss that requires periodic replacement pH probes age through normal use a...

Страница 9: ...le Replacement Replace your calibration solution within 2 months of opening for continued accuracy 9 When Heavily Contaminated Heat a diluted Potassium Chloride KCl solution to 60 80 degrees Celsius Place the tip of the probe into the heated solution and let sit for an hour Remove the probe from the solution and let it cool to room temperature Then clean with a toothbrush Rinse with pH 4 solution ...

Страница 10: ...Specifications Abilities 10 Operational Environment 20 to 80C 4 to 176 F ...

Страница 11: ...ration guide 6 Ensure the PPM probe is fully plugged into the meter 7 Replace your Nutrient probe 11 Troubleshooting Guide Problem No Display Follow the steps below until a solution is reached 1 Check that the Growboss is connected to the 5dcv 0 5 amp power supply 2 Check that the Power supply is connected to the wall outlet 3 Verify that the outlet has power by connecting another device such as a...

Страница 12: ...recalibrate as per procedures 3 Leave probe in solution for at least 10 minutes 4 Replace probe s 12 Problem Won t Hold Calibration Follow the steps below until a solution is reached 1 Power off the unit Unplug probe s and carefully plug back in ensuring connections are tight 2 Replace calibration solution and recalibrate as per procedures 3 Leave probe in solution for at least 10 minutes 4 Replac...

Страница 13: ... directly by Nutradip All warranty claims must be submitted online at www futureharvest com warranty 13 Warranty Coverage The Nutradip GrowBoss is covered by three separate warranty programs The 12 month warranty coverage protects the product from material workmanship defects The Accidental Damage Coverage plan protects product from damage caused by user error or negligence The theft coverage plan...

Страница 14: ...due to defective materials or workmanship existing at the time of purchase If any part is no longer available or out of maufacture Future Harvest will replace it with a functionally equivalent replacement part Future Harvest shall not be liable for costs of repair or replacement of a product incurred as a result of 1 Normal wear and tear 2 Accidental damage faults caused by negligent use or care n...

Страница 15: ...NTIES GUARANTEES AND REMEDIES WHETHER ORAL OR WRITTEN EXPRESS OR IMPLIED ANY IMPLIED WARRANTIES THAT MAY BE IMPOSED BY LAW INCLUDING WITHUT LIMITATION MERCHANTABILITY ARE LIMITED IN DURATION TO THE PERIOD OF THIS LIMITED GUARANTEE TO THE MAXIMUM EXTENT PERMITTED BY LAW FUTURE HARVEST IS NOT RESPONSIBLE FOR AND CONSUMER RELEASES FUTURE HARVEST FROM SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULT...

Страница 16: ......

Отзывы: