4
Type
NE214
Rated power supply
AC100V 50/60Hz, AC120V 50/60Hz, AC230V 50/60Hz
Resonance frequency
28~32kHz
Maximum output
17W
Power supply input
50VA
Lighting
White LED (Optic Handpiece (VS-LED-HPSC) only)
Activate Voltage
3.5V
Consumption current
0.1A (Typ. 3.5V)
Dimensions
W268 x D230 x H103(mm)
Weight
3.1kg
Fuse
100/120V Specification...T2AL 250V
230V Specification...T1AL 250V
Use environment
Temperature
0 - 40 ˚C*
Humidity
30 - 75 %**
Atmospheric pressure
70 - 106 kPa
Transportation and Storage
environment
Temperature
-10 - 60 ˚C
Humidity
10 - 85 %**
Atmospheric pressure
50 - 106 kPa
* No liquid freezing allowed
** No condensing allowed
2. Specifications
Principle of Operation
A sinusoidal electrical signal, at ultrasonic frequency, is delivered by the generator. This signal is applied to the
piezoelectric ceramic located inside the transducer. Piezoelectric ceramic converts this signal into mechanical
vibrations. These vibrations are at the same ultrasonic frequency as the electrical signal. The mechanical vibrations
are propagated towards the distal end of the transducer. The insert, which is attached at the distal end of the
transducer, vibrates at ultrasonic frequency and makes it possible to achieve the aimed purpose.