12
Wrench
Fig. 10
Tighten
Loosen
Ring and
attaching screw
Ring and
attaching screw
9. Utrasonic cleaning (X57L,X57)
To remove the paste that against into the Handpiece, perform ultrasonic cleaning day by day.
1) Loosen the ring and take off the ring and attaching screw.(Fig.10)
2) Perform the ultrasonic cleaning for 5 minutes ONLY for the attaching screw.
3) Air blow and remove the water of the head. Than lubricate it by PANA SPRAY Plus after the opposite
procedure of 1.
4) Autoclave the handpiece. (6. Sterilization)
CAUTION
• DO NOT use ethanol or cleansing fluid for the cleaning. It may cause tarnish or fret.
• Repeat this lubrication until you see no paste into the handpiece comes out.